Eschmann · Model T10 -m
Powered operating table for general and specialist surgery. 250kg maximum patient weight static, 200kg mobile. Full length radiolucent top with integral x-ray cassette tunnel accessible from either end. 753-1053mm height range with mattresses. 600mm wide table top (across sidebars) - optional width extenders available. Corded handset controlling Trendelenburg, reverse Trendelenburg, lateral tilt, flexion/extension and height. Offset table-top break and offset slimline column for superior imaging and access. Column covers in stainless steel. Battery powered, with facility for connection to mains electricity for immediate use. Low battery warning and battery exhaustion protection. Foot pedal on base to control brakes, wheels and castors for stability, steering and straight-line and 360 degree mobility. Can be accessed from head end or either side of table. Head and leg sections are removable and interchangeable, with an auto-locking mechanism, and can be adjusted with one hand. Covers to the top and base are made of a scratch resistant and colourful acrylic capped ABS mouldings to minimise potential sites for cross contamination and provide a wash down surface to the tabletop. Mattress are moulded and anti-static with no seams so as to remove potential sites on the mattresses where body fluids such as blood can congregate and also to ensure ease of cleaning after each procedure or operation. Special one-piece mattress covering break position (long and short trunk sections) to combat fluid ingress and to provide superior patient support. Standard configuration has 50mm deep mattress, with 75mm mattress option. Optional colour co-ordinated castor covers available. Option of table with divided legs available.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Height, tilt, Trendelenburg, back or leg section does not respond to the hand control / pendant.
Root cause: Hand control fault, an electro-hydraulic actuator, a control-board fault, or a low battery (mobile column).
Symptoms: Table settles under load after positioning.
Root cause: Hydraulic valve internal leak or an actuator backdrive.
Symptoms: Short run time, table dies mid-procedure, battery alarm.
Root cause: Aged battery pack or a charger fault.
Symptoms: Intermittent or dead hand control.
Root cause: Worn hand control or a damaged cable.
Symptoms: Segments loose, accessory rails will not clamp, pads torn, radiolucent top damaged.
Root cause: Worn segment latches, rail clamps, or upholstery.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Electro-hydraulic actuator | Major service part | $$$ | Difficult |
| Hydraulic control valve block | Major service part | $$$ | Difficult |
| Battery pack (mobile column) | Periodic replacement | $$ | Fairly easy |
| Hand control / pendant | Periodic replacement | $$ | Fairly easy |
| Table-top segment pads / rail clamps | Major service part | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio