Diatron · Model Abacus 3
Color LCD, the large display allows rapid and accurate patient data entry and monitoring of the results with large histogram Sample rotor, ensuring safe and easy sampling The instrument features a sophisticated but easy-to-use, logically constructed multilingual operating menu. It provides 6-level QC, L-J graphs and self-diagnostic functions to let you monitor reliability and accuracy All Abacus models can be interfaced with your host computer for further processing of the measured data. Results can also be retrieved on the built-in roll paper printer Abacus 3 provides 18-parameters, a complete CBC including a 3 part WBC differential High throughput of 60 samples per hour, results within 60 seconds The fluidic system is automatically cleaned from dust particles when turning on the analyzer. High voltage pulses after each measuring cycle keep the apertures free from clogging. The instrument automatically cleans the sampling tip inside and outside after blood aspiration to minimize the risk of carryover and infection. Abacus 3 is almost maintenance free resulting in minimal disruption of operation Clinical environment by utilizing industry standard electronic and mechanic components.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Flagged or implausible CBC counts, voteouts, clog alarms, poor precision.
Root cause: Debris or a clot partially blocking a counting aperture or the flow cell.
Symptoms: Short samples, aspiration errors, cap-piercing failures, carryover.
Root cause: Worn or bent sample probe / closed-vial piercer, or a worn shear / segment valve.
Symptoms: WBC, HGB or diff results shifted, high background, reagent alarms.
Root cause: Reagent depleted or wrong lot, or a pinch-valve / tubing fault.
Symptoms: Diff scattergrams abnormal, missing populations, flow-cell or laser alarm.
Root cause: Contaminated or aged flow cell, or the laser / optics.
Symptoms: Multiple parameter errors, priming failures.
Root cause: Pump or regulator drift.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Sample probe / closed-vial piercer | Periodic replacement | $$ | Moderate |
| Shear / segment valve | Periodic replacement | $$$ | Difficult |
| Pinch-valve / peristaltic tubing | Frequent wear item | $ | Moderate |
| Flow cell | Periodic replacement | $$$ | Difficult |
| System reagents / controls | Consumable | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.