Bausch & Lomb · Model MILLENNIUM
MODULAR SYSTEM FOR CATARACT AND VITREOUS REMOVAL The Millennium microsurgical system delivers on the promise of a platform with innovative technology. For the latest advancement in fluidics control for the cataract surgeon, Bausch & Lomb Surgical offers Millennium™ Advanced Flow System (AFS) and Custom Control Software (CCS) which take you beyond cold phaco with a new hyper-cool level of control with your choice of either a venturi or peristaltic system. The Dual Linear™ control feature allows for simultaneous control of either flow or vacuum, and ultrasound power. Additionally, surgeons can use both flow and vacuum response in the same procedure. The powerful Quad-crystal ultrasound handpiece delivers excellent cutting with smooth efficiency. Programming options allow for virtually unlimited parameters of storage and surgeon-controlled mode switching. With its modular design, the Millennium system integrates innovation for today and the future.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Weak ultrasonic power, poor followability, vacuum will not build or holds poorly, surge.
Root cause: Worn phaco handpiece / tip, a cracked or worn tubing / cassette, or a pump / vacuum-sensor fault.
Symptoms: Handpiece will not tune, overheats, no irrigation / aspiration, cutter not cutting (vit).
Root cause: Worn or damaged handpiece, or its connector / cable.
Symptoms: Cassette not recognised, priming / tuning fails, fluid leaks at the cassette.
Root cause: Cassette-detection switch, the peristaltic / Venturi pump, or fluidics seals.
Symptoms: Pedal positions not registering, wireless pedal not pairing.
Root cause: Worn footpedal or its cable / wireless module.
Symptoms: Cut rate low or erratic, cutter will not drive.
Root cause: Internal pneumatic valve or a supply-pressure fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Phaco / I-A / vitrectomy handpiece | Periodic replacement | $$$ | Fairly easy |
| Peristaltic / Venturi pump components | Periodic replacement | $$$ | Difficult |
| Cassette-detection switch / fluidics seals | Periodic replacement | $$ | Moderate |
| Footpedal / wireless module | Periodic replacement | $$ | Fairly easy |
| Pneumatic valve (vitrectomy) | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
The AC power cords used on these products are subject to an FDA Alert -- Safety Investigation of Certain Medical Device Power Cords: Initial Communication, Date Issued: October 19, 2009.
Users may incorrectly assemble the plastic needle wrench to the phacoemulsification needle resulting in the generation of plastic particulate
The directions for use are printed with the incorrect symbols to indicate the product is packaged as sterile and nonreusable
The cutter probe tip breaks during use.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.