
Axia Surgical · Model Flex 3000P
Modern and easy to clean stainless steel base. Generous imaging area. Reverse orientation mode. Suggested Featured Accessories. Advanced Intuitive Table. Multiple surgical applications.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Height, tilt, Trendelenburg, back or leg section does not respond to the hand control / pendant.
Root cause: Hand control fault, an electro-hydraulic actuator, a control-board fault, or a low battery (mobile column).
Symptoms: Table settles under load after positioning.
Root cause: Hydraulic valve internal leak or an actuator backdrive.
Symptoms: Short run time, table dies mid-procedure, battery alarm.
Root cause: Aged battery pack or a charger fault.
Symptoms: Intermittent or dead hand control.
Root cause: Worn hand control or a damaged cable.
Symptoms: Segments loose, accessory rails will not clamp, pads torn, radiolucent top damaged.
Root cause: Worn segment latches, rail clamps, or upholstery.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Electro-hydraulic actuator | Major service part | $$$ | Difficult |
| Hydraulic control valve block | Major service part | $$$ | Difficult |
| Battery pack (mobile column) | Periodic replacement | $$ | Fairly easy |
| Hand control / pendant | Periodic replacement | $$ | Fairly easy |
| Table-top segment pads / rail clamps | Major service part | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio