Visaris Americas · Model Vision V
Motorised vertical and horizontal detector tracking. Sync mode and auto light-on when stand is in motion. Detachable wireless tube-side console for remote system control. Scalable generator power and high power tube options. Motorised collimation controls on bucky stand. Auto collimation on worklist with SID auto-adjustment. DAP meter with system (DICOM) integration. Exposure stop when tube not synchronised with detector HIGH PERFORMANCE IMAGING. Vision V incorporates a powerful digital radiography control and acquisition system (Avanse DR) with flexible single or multi detector configuration. System console with full DR functionality from patient search/entry, direct generator control, fast image acquisition and processing to DICOM image archiving and export offers unparalleled ergonomy and efficiency of the examination process. Intelligent, automated, procedure specific generator, collimation, stand and image processing organ program settings make Avanse DR optimal for high patient throughput with exceptional image quality. Imaging console seamlessly integrates with portable tube-side console, radiography stand and smart buckys on the Vision V system and can also be enhanced with a range of Digital Radiology components such as VisarisPACS and Visaris Diagon diagnostic workstation software to turn it into a complete digital radiology department. Flexible detector configurations: fixed, portable, wireless 35×43, 43×43. High sensitivity detectors for optimal image quality. Integrated smart bucky with detector charging, rotation and removable grid. Detector orientation, grid, SID and stand position display on system console. Automatic exposure adjustment with changing SID and grid. Integrated Avanse service and remote support tools.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Panel not detected, images not received, wireless / tethered link drops mid-exam.
Root cause: Detector battery (wireless), Wi-Fi / tether cable, a docking / connector fault, or detector electronics.
Symptoms: Fixed-pattern artefacts, non-uniformity, dead pixels / lines, calibration overdue or failing.
Root cause: Detector gain / offset calibration, a bad pixel map, or physical detector (drop) damage.
Symptoms: Exposure inhibited, kV / mAs error, no exposure captured, AEC over/under.
Root cause: HV generator, tube, filament supply, or an AEC / interface fault.
Symptoms: Tube column, wall stand or table bucky will not track, detent, or auto-position.
Root cause: Drive motor, brake, encoder, or the auto-positioning interface.
Symptoms: Images will not display, process or send; worklist missing; slow processing.
Root cause: Workstation PC, software, or the DICOM / network configuration.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Flat-panel detector (exchange) | Major service part | $$$$ | Moderate |
| Detector battery (wireless) | Periodic replacement | $$ | Easy |
| Tether cable / access point | Periodic replacement | $$ | Fairly easy |
| Positioner drive / brake / encoder | Major service part | $$$ | Difficult |
| Acquisition workstation PC | Major service part | $$$ | Moderate |
| AEC ion chamber | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.