Stryker · Model FluidSafe
Key advantages of the FluidSafe fluid management system include. Pre-set pressure and flow rates. Adjustable audible and visual deficit warnings. Automatic detection of fluid bag exchanges. Customized flow rate up to 500 ml/min and pressure up to 200mmHg. Bright LCD pump display with intuitive controls.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Set pressure sags or spikes, poor distension, pressure alarms.
Root cause: Inflow pump rollers, a pressure-sensor fault, or a tubing-set fault.
Symptoms: Deficit reads implausibly or will not tare, alarm at the wrong volume.
Root cause: Weigh scale / load cell out of calibration, or bags not hung freely.
Symptoms: Poor clearance of the cavity, fluid backs up.
Root cause: Vacuum / outflow pump, canister seal, or tubing occlusion.
Symptoms: System will not prime or recognise the disposable set.
Root cause: Cassette mis-seated or a set-detect sensor fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Recalibrate deficit accuracy after replacement.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Inflow pump roller assembly | Periodic replacement | $$$ | Difficult |
| Weigh scale / load cell (FluidSafe) Recalibrate deficit accuracy after replacement. | Major service part | $$$ | Difficult |
| Vacuum / outflow pump | Periodic replacement | $$$ | Difficult |
| Single-use tubing set | Consumable | $ | Easy |
| Pressure transducer | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio