Allen Medical · Model Thermedx® FluidSmart™
Simple Set-Up and Procedure Selection. On-Demand Fluid Warming (Up to 40º C). Adjustable Pressure Settings (Non-Pulsatile Flow). Flow Rate- Automatic flow rate adjustment provides constant pressure for even distention. Continual Irrigation Capacity. Convenient Operation. Substantial Safety Features. Fluid Deficit Monitoring (Deficit Alarm up to 2,500 ml).
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Set pressure sags or overshoots, poor distension, pressure alarms.
Root cause: Inflow pump rollers, a pressure-sensor fault, or a tubing-set fault.
Symptoms: Return fluid cool, warming not reaching up to 40 C.
Root cause: Heater cartridge or its temperature control.
Symptoms: Deficit reads implausibly or will not tare; alarm at the wrong volume.
Root cause: Weigh scale / load cell out of calibration or bags not hung freely.
Symptoms: System will not prime or recognise the disposable set.
Root cause: Cassette mis-seated or a set-detect sensor fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Recalibrate deficit accuracy after replacement.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Inflow pump roller assembly | Periodic replacement | $$$ | Difficult |
| Fluid-warming heater cartridge | Periodic replacement | $$$ | Moderate |
| Weigh scale / load cell Recalibrate deficit accuracy after replacement. | Major service part | $$$ | Difficult |
| Single-use tubing set | Consumable | $ | Easy |
| Pressure transducer | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio