Fluid management system for hysteroscopic procedures. With a touchscreen user interface, improved operation settings and rapid response to changes in intrauterine pressure, the HysteroLux™ system offers a next generation solution for hysteroscopic procedures. The HysteroLux™ system enables continuous, real-time monitoring of fluid deficit when its scale is in use. And it reduces the risk of fluid overload by continuously monitoring fluid volume deficit with ±6 percent accuracy.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Pressure sags or spikes vs the set point, poor distension, pressure alarms.
Root cause: Inflow peristaltic pump rollers, pressure transducer drift, or a tubing-set fault.
Symptoms: Deficit reads implausibly, jumps, or will not tare; outside the +/-6% spec.
Root cause: Weigh scale / load cell out of calibration, bags not hung freely, or a tare error.
Symptoms: Poor clearance of the cavity, fluid backs up.
Root cause: Vacuum pump, canister seal, or outflow tubing occlusion.
Symptoms: System will not recognise or prime the disposable set.
Root cause: Cassette mis-seated or a set-detect sensor fault.
Symptoms: Unresponsive or erratic touchscreen.
Root cause: Display or controller fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Recalibrate deficit accuracy after replacement.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Inflow peristaltic pump roller assembly | Periodic replacement | $$$ | Difficult |
| Weigh scale / load cell Recalibrate deficit accuracy after replacement. | Major service part | $$$ | Difficult |
| Vacuum pump | Periodic replacement | $$$ | Difficult |
| Single-use tubing set | Consumable | $ | Easy |
| Pressure transducer | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.