Hologic · Model Fluent
Simplified setup and operation Advanced technology to increase clinical confidence Streamlined workflow for the entire OR team.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: "Supply bag missing/empty", "Waste bag missing/full", "Check fluid/waste bag hooks" alarms; deficit calculation wrong.
Root cause: Bag not hung on both hooks, a jostled or empty bag, an object on the hooks, or a load-cell reading error.
Symptoms: Poor visualisation, cavity not clearing, low return flow.
Root cause: Outflow tubing not connected to the hysteroscope or outflow port, buttocks-drape tubing off, or a blocked outflow path.
Symptoms: "In-FloPak not detected" at setup, pump will not start.
Root cause: Cassette not seated fully, or worn cassette-detect sensor / pump interface.
Symptoms: Set pressure not maintained, visibility issues, pump over/under-correcting.
Root cause: Pressure sensor drift, worn peristaltic pump rollers, or a fluid-bag clamp left closed.
Symptoms: Persistent hook alarms after correct bag hanging, deficit numbers implausible.
Root cause: Bent or contaminated hook, or a failed load cell needing calibration.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Calibrate the fluid-deficit measurement after replacement.
Single-procedure consumable.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Supply / waste load cell Calibrate the fluid-deficit measurement after replacement. | Major service part | $$$ | Difficult |
| Bag hook assembly | Major service part | $$ | Moderate |
| Peristaltic pump roller head | Periodic replacement | $$$ | Moderate |
| Cassette-detect sensor / pump interface | Major service part | $$ | Moderate |
| In-FloPak disposable cassette and tubing set Single-procedure consumable. | Consumable | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.