ST. JUDE MEDICAL, INC. · Model Model 3077
Designed for safe and reliable temporary stimulation of the heart in cases of rhythm disturbances and conduction defects An extensive single-chamber feature set, including mode choices, a wide base rate range, adjustable amplitude and sensitivity parameters, and two modes of high-rate pacing Up to 12 volts of output available per channel Standard 9 volt lithium or alkaline batteries are used, and the device features both visual and audible battery life indicators Large, simple dial Small size and lightweight design Runaway protection limits the device to a maximum rate of 200 ppm in the unlikely event of circuit malfunction
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Charge fails or sluggish, low-energy delivery.
Root cause: Capacitor aging / leakage.
Symptoms: Short charge-hold, low-battery on readiness test.
Root cause: Aged battery pack.
Symptoms: No shock delivered, pad error.
Root cause: Worn paddles or expired pads.
Symptoms: Leakage alarm, self-test fail.
Root cause: Failed isolation components.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Expiration-dated consumable.
Critical - keep charged spares.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Defibrillation electrodes / pads Expiration-dated consumable. | Consumable | $$ | Easy |
| Charge relay / contactor | Periodic replacement | $$$ | Difficult |
| Patient interface cable | Periodic replacement | $$ | Fairly easy |
| Rechargeable battery pack Critical - keep charged spares. | Periodic replacement | $$$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →