Medtronic · Model 5388
Simply adjusting the top three dials provides effective pacing in most clinical situations. Changing the pacing rate automatically adjusts other dual-chamber temporary pacing parameters. Pediatric patients may need a pacing rate up to 200 ppm. A patient's high stimulation threshold may require a ventricular output of 25 mA. For managing atrial tachyarrhythmias, a rapid atrial pacing rate up to 800 ppm is possible. The patient cable system uses protected electrical connections to increase patient safety. The Model 5388 incorporates the quality and durability users expect from Medtronic temporary pacemakers.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Pacing stimuli not followed by a paced beat, capture lost intermittently.
Root cause: Output (mA) set below threshold, a loose / oxidised connection at the terminal block or the bridging / patient cable, or a lead-contact issue at the patient.
Symptoms: Pacing spikes fall on intrinsic beats (undersensing) or pacing is inhibited inappropriately (oversensing).
Root cause: Sensitivity (mV) set wrong, EMI, or a cable / connection fault.
Symptoms: Low-battery indicator, or the device warns of imminent depletion.
Root cause: 9 V battery low - replace immediately; keep a spare with the device.
Symptoms: A warning message when adjusting RATE or a rate-dependent parameter, the value will not go further.
Root cause: The device enforces minimum V-A intervals (70 ms in DOO, 180 ms in DVI) - the requested combination is not allowed.
Symptoms: Intermittent output or sensing when the cable is moved.
Root cause: Worn terminal block or a damaged patient / bridging cable.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Replace on any low-battery indication; always keep a spare with the device.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| 9 V battery Replace on any low-battery indication; always keep a spare with the device. | Consumable | $ | Easy |
| Patient / bridging cable | Periodic replacement | $$ | Easy |
| Terminal block | Major service part | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.