Spacelabs · Model 90343
SPACELABS 90343 The Spacelabs Medical Digital Telemetry system provides continuous electrocardiographic (ECG) monitoring to detect ST segment changes, abnormal cardiac rhythms, including asystole, ventricular fibrillation, and ventricular tachycardia. When the system is used with the multiparameter digital telemetry transmitter (90343) and the Ambulatory Blood Pressure (ABP) monitor (90217), monitoring of ECG signals is enhanced by the availability of continuous or episodic pulse oximetry (SpO2) measurements and/or episodic noninvasive blood pressure (NIBP) measurements.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: A patient's waveform drops out or is noisy at the central station, coverage gaps on the floor.
Root cause: Weak or aged transmitter, depleted transmitter batteries, antenna / receiver-system fault, or RF interference / dead spots.
Symptoms: A whole bank of channels drops at the central station, a housing shows offline.
Root cause: SDLC cabling between J100-J400, a receiver-module fault, or a power fault in the housing.
Symptoms: Overheating, intermittent channel loss; no airflow felt in a module bay.
Root cause: Failed housing cooling fan.
Symptoms: Noisy or absent ECG from one transmitter regardless of location.
Root cause: Worn lead set or the transmitter ECG input.
Symptoms: Central monitor will not display telemetry, freezes, or will not admit a patient.
Root cause: Central-station PC or software fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Telemetry transmitter | Major service part | $$$ | Easy |
| Transmitter batteries / lead set | Consumable | $ | Easy |
| Receiver module | Major service part | $$$ | Moderate |
| Receiver-housing cooling fan | Major service part | $ | Moderate |
| SDLC / antenna cabling | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.