Key Features Integrated hemodynamic and airway gas monitor designed for operating rooms and recovery areas Intuitive Datex-Ohmeda user interface with anesthesia-specific menu navigation and alarm management Compact, space-saving design for environments where space is limited Large numeric values and waveforms for excellent visibility from a distance Built-in backup battery for uninterrupted operation during unexpected power loss Integrated mounting system supports quick installation on optional wall mounts, roll stands, and anesthesia machine mounting arms.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: CO2, O2, N2O or agent values off vs a reference, calibration prompts, agent ID wrong.
Root cause: Gas-bench aging or a missed calibration, or contamination in the sampling path.
Symptoms: Occlusion or sample-flow alarm, no or low sample flow, flat gas waveforms.
Root cause: Full or expired D-fend water trap, a blocked / kinked sampling line, or a blocked occlusion filter.
Symptoms: Slow gas response, low sample flow that a new trap and line do not fix.
Root cause: Worn internal sample pump.
Symptoms: Noisy or absent trace, no reading, NIBP times out.
Root cause: Worn patient cable / sensor, or a monitor input fault.
Symptoms: NIBP times out, cuff will not inflate / hold, readings implausible.
Root cause: Internal NIBP pump or valve, a cuff / hose leak, or a pressure-transducer fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Also replace the occlusion filter, zero absorber, Nafion tubing and OM reference filter at PM.
Use only Datex-Ohmeda sampling lines.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| D-fend water trap and O-ring Also replace the occlusion filter, zero absorber, Nafion tubing and OM reference filter at PM. | Consumable | $ | Easy |
| Sampling line Use only Datex-Ohmeda sampling lines. | Consumable | $ | Easy |
| Internal sample pump | Periodic replacement | $$$ | Difficult |
| O2 sensor | Frequent wear item | $$ | Fairly easy |
| Patient cable set / NIBP cuff and hose | Frequent wear item | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.