Proven safety and reliability State of the art hardware and software technology Fully flexible and configurable modular concept Functional and easy to use
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Poor gas exchange or temperature control.
Root cause: Oxygenator membrane fouling or heat-exchanger leak.
Symptoms: Inaccurate flow, unusual noise from the pump head.
Root cause: Raceway wear or centrifugal rotor bearing wear.
Symptoms: False or missed air-in-line/level alarm.
Root cause: Ultrasonic sensor drift or dirty optics.
Symptoms: Console freeze or loss of readouts mid-case.
Root cause: Main control board or display module fault.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Safety-critical; verify at PM.
Sterilizable disposable raceway; console head inspected at PM.
Single-use per case.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Level / bubble sensor Safety-critical; verify at PM. | Major service part | $$$ | Moderate |
| Console main board | Major service part | $$$$ | Expert-level |
| Pump head / raceway Sterilizable disposable raceway; console head inspected at PM. | Major service part | $$$$ | Difficult |
| Oxygenator (disposable) Single-use per case. | Consumable | $$$$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
Identification of excess plastic on the tip of the cannula.
Sorin Group is recalling certain lots of Edwards Lifesciences Suction Wand Model S099B due to presence of plastic particles inside the pouch.
Touch screen may be unresponsive, inhibiting user input.
Customer complaints of a leak between the blood and water compartments of the heat exchanger used in the oxygenators due to prolonged contact of the heat exchangers with sodium chloride causing micro-holes.
Touch screen may become unresponsive, inhibiting user input.
Defective product packaging may compromise product sterility.
Tubing set packaging seal between lid and tray may not be adequately sealed, compromosing sterility.
Vessel clamping device may break during use.
A weakness in the plastic packaging tray could cause the tray to crack; compromising product sterility.
Under certain conditions, venous reservoir on oxygenator may allow air into the system.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.