Terumo · Model Sarns 8000
Modular design Four or five pump base Roller or centrifugal pumps Arterial and cardioplegia monitors for temperature, pressures and timers Battery module Electronic venous occluder Pulse mode Communications module for data acquisition.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: A pump will not run at set flow, speed unstable, occlusion out of adjustment, flow-probe error.
Root cause: Worn roller-pump raceway / rollers, a centrifugal drive-motor / coupling fault, or a flow-probe fault.
Symptoms: A pump module unresponsive, display errors, module not recognised by the console.
Root cause: Module electronics or the console-to-module link.
Symptoms: Short back-up run time, will not transfer cleanly on a power interruption.
Root cause: Aged battery / UPS pack.
Symptoms: Pressure, reservoir-level or bubble-detector alarms implausible or not triggering.
Root cause: Transducer, level-sensor, or bubble-detector fault.
Symptoms: Blood / cardioplegia temperature not tracking, heater-cooler alarm, flow fault.
Root cause: Heater-cooler pump / refrigeration or a temperature-probe fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Roller-pump raceway / roller kit | Periodic replacement | $$$ | Difficult |
| Centrifugal drive motor / coupling | Major service part | $$$ | Difficult |
| Battery / UPS pack | Periodic replacement | $$$ | Moderate |
| Pressure / level / bubble sensors | Periodic replacement | $$ | Moderate |
| Flow probe | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Terumo Cardiovascular Systems (Terumo CVS) has received multiple reports since 2007 of various malfunctions of the power distribution and battery back-up systems for the Sarns Modular Perfusion System 8000. Very few of the reports occurred during cardiopulmonary bypass; the majority of the malfunctions occurred during set-up or were reported during routine preventative maintenance. The malfunction
Terumo Cardiovascular Systems (Terumo CVS) has received multiple reports since 2007 of various malfunctions of the arterial and cardioplegia monitors for the Sarns Modular Perfusion System 8000. The malfunctions can be broadly categorized as follows monitor power malfunctions, inability to measure pressure and inaccurate and/or false pressure alarms. If the malfunction of the arterial or cardio
During a retrospective review of complaint events, TCVS identified on 06/15/12 that three (3) Sarns" System 8000 bases were manufactured and distributed from 11/13/06-11/29/06 with an incorrect incorrect amperage rating identified on the name plate label. The mislabeling was identified in January 2007 a set up process of the automark label printer. All units were corrected in the field in Marc
Terumo received multiple reports of malfunctions of the Cardioplegia and arterial monitors for the Sarns Modular Perfusion System 8000. Upon review of complaint investigation, the malfunctions are caused by or could have been caused by soldering anomalies on the circuit board of the monitor. The compromised solder joints could cause intermittent or loss of display, or could cause monitor reset, po
Terumo Cardiovascular Systems is aware on one instance in which an incorrectly labeled printed circuit board for the Sarns Modular Perfusion System 8000 was shipped to a single customer. The affected board mislabeled two connection between the board and the transformer.
The Cardioplegia Monitor for the Sarns Modular Perfusion System 8000 may fail to track cardioplegia delivery volume.
The system 8000 roller pumps may fail resulting in a persistent pump stop.
Terumo Cardiovascular Systems (Terumo CVS) received complaints from our Affiliate, Terumo Dubai that the safety monitors were tripping off. Terumo CVS' investigation determined that 8 monitors were manufactured with an underrated circuit breaker due to an error where an associate pulled an underrated circuit breaker from an incorrect inventory location. Since this time, inventory locations hav
False Alarms: The air sensor may malfunction and trigger false alarms, and may continue to alarm thus preventing the device from being reset.
The hand crank may crack in the bearing/shaft area due to a manufacturing defect. (Note: Some of the hand cranks subject to this recall were shipped as replacements for hand cranks recalled in the Summer of 2006.)
Roller Pump: Due to a circuit board malfunction the roller pumps may stop during use without displaying an error message, may stop during use with an overspeed error posted on the local display, or may fail to start/restart resulting in the pump(s) being non-functional.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.