Smiths Medical · Model Level 1 H-1200
Rigid pressure chambers accommodate standard blood and crystalloid bags providing a constant 300mmHg pressure for rapid infusion. On/Off toggle switch used to quickly and easily pressurize chambers. Aluminum heat exchanger, that transfers heat 1000 times faster than plastic, and the counter current 42°C circulating water bath, ensure patients receive normothermic blood and I.V. fluids. Additional level of safety with the integrated Air Detector/Clamp that detects the presence of air in blood and crystalloid, alerts clinicians, and automatically stops the flow and allows quick removal of air without disconnecting from the patient or changing disposables.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Output below target, low-temperature alarm, slow warm-up.
Root cause: Heater plate(s) or the heater-control circuit, or a temperature-sensor fault.
Symptoms: The unit disables heating with an over-temperature condition (protects against haemolysis).
Root cause: Sensor or control fault driving the heater high, or an extreme flow-rate change.
Symptoms: Air-in-line alarm, flow stops.
Root cause: Air in the set, an improperly primed set, or an air-detector fault.
Symptoms: Set not detected, will not clamp / seat, leaks at the warming chamber.
Root cause: Worn set-detection switch or chamber fitting.
Symptoms: Cannot reach set flow, pressure alarm.
Root cause: Pump / roller fault or an occlusion.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Heater plate(s) / heater-control assembly | Periodic replacement | $$$ | Moderate |
| Temperature sensor | Major service part | $$ | Moderate |
| Air-detector / bubble sensor | Major service part | $$ | Moderate |
| Set-detection switch / chamber fitting | Periodic replacement | $$ | Fairly easy |
| Disposable warming set | Consumable | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
IV Fluid Administration Set may have a blockage of the recirculating warming fluid channel within the disposable administration set, causing an alarm condition of over-temperature during priming.