3M Health Care · Model 24500
• Safely and effectively warms fluids from KVO to 30,000 mL/hr • Quickly reacts to changes in flow rate • Uses economical, easy-to-use disposable warming sets • Meets all American Association of Blood Bank (AABB) guidelines for warming blood • Quick and simple maintenance • Highly conductive warming plates maximize heat transfer • No excess heat is retained in the system, so flow changes will not overheat fluid • Low-resistance flow path is engineered for optimum performance • Warming plates are ideally spaced for optimal surface area contact • Built-in safety features provide added protection • Operating set point at 41°C • Audible and visible system alarms if fluid temperature exceeds 42°C • Audible and visible under-temperature alarms • Secondary alarm system provides fail-safe backup
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: No heat, or overheat alarm.
Root cause: Heater element or thermal cutoff failure.
Symptoms: Delivered fluid temperature out of set range.
Root cause: Inline/probe sensor aging or miscalibration.
Symptoms: False or missed air-in-line alerts.
Root cause: Air detector sensor drift or dirty optics.
Symptoms: Fluid leak, or won't prime.
Root cause: Worn cassette/tubing seat or cracked fitting.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Single-use consumable.
Consumable in some designs.
Safety-critical; verify at PM.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Disposable tubing / cassette set Single-use consumable. | Consumable | $ | Easy |
| Heating element / plate | Periodic replacement | $$$ | Moderate |
| Temperature sensor / probe Consumable in some designs. | Periodic replacement | $$ | Fairly easy |
| Air-in-line sensor Safety-critical; verify at PM. | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →