Barkey · Model Autocontrol XPT
Prevention of hypothermia through warmed infusions and transfusions. Designed for any standard I.V. set. Adjustable temperature range between +33°C to +43°C. No additional disposable required. Language selection. Can be applied to the return line of dialysis fluids. Option to connect the warming device autoline 4R XPT to the control unit autocontrol 3 XPT. Autocontrol 3 XPT automatically recognizesthe connected warming device autoline 4R XPT. Displays preset and actual temperature of the warming devices. Use of biocompatible materials. Visual and acoustic alarm in case of error. Safety through three independent temperature sensors. Turnable stand and standard rail fixing. Overtemperature alarm (1°C above set point).
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: The yellow indicator lights and the display shows a numeric error code for the last fault.
Root cause: A specific condition indexed in the manual's error-code list - a heater-control-line / plate fault, a temperature-sensor fault, or a control-unit fault.
Symptoms: Plasma / product not thawed / warmed to target in the expected time.
Root cause: Heater plates or the heater-control lines, or a temperature-sensor fault.
Symptoms: One heater plate not warming, uneven thaw.
Root cause: A failed heater plate or a broken heater-control line.
Symptoms: The unit disables heating with an over-temperature condition.
Root cause: Sensor or control fault driving the heater high.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Heater plate(s) | Periodic replacement | $$$ | Moderate |
| Heater-control lines | Periodic replacement | $$ | Fairly easy |
| Temperature sensor | Major service part | $$ | Moderate |
| Control unit / board | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.