• Prevention of hypothermia through warmed infusions and transfusions • Designed for any standard I.V. set • Adjustable temperature range between +33°C to +43°C • No additional disposable required • Language selection • Can be applied to the return line of dialysis fluids • Option to connect the warming device autoline 4R XPT to the control unit autocontrol 3 XPT • Autocontrol 3 XPT automatically recognizesthe connected warming device autoline 4R XPT • Displays preset and actual temperature of the warming devices • Use of biocompatible materials • Visual and acoustic alarm in case of error • Safety through three independent temperature sensors • Turnable stand and standard rail fixing • Overtemperature alarm (1°C above set point)
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: No heat, or overheat alarm.
Root cause: Heater element or thermal cutoff failure.
Symptoms: Delivered fluid temperature out of set range.
Root cause: Inline/probe sensor aging or miscalibration.
Symptoms: False or missed air-in-line alerts.
Root cause: Air detector sensor drift or dirty optics.
Symptoms: Fluid leak, or won't prime.
Root cause: Worn cassette/tubing seat or cracked fitting.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Single-use consumable.
Consumable in some designs.
Safety-critical; verify at PM.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Disposable tubing / cassette set Single-use consumable. | Consumable | $ | Easy |
| Heating element / plate | Periodic replacement | $$$ | Moderate |
| Temperature sensor / probe Consumable in some designs. | Periodic replacement | $$ | Fairly easy |
| Air-in-line sensor Safety-critical; verify at PM. | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.