ARTIS icono ceiling It’s easy to see why interventionalists are talking about the new ceiling-mounted C-arm from Siemens Healthineers. It delivers movement flexibility and faster rotational speed not seen in other systems, for shorter acquisition time in 3D. The unique imaging chain features self-adjusting algorithms that regulate the image impression to achieve constant image quality at the lowest achievable dose that can be tailored to each user. ARTIS icono biplane ARTIS icono biplane is the ideal solution for a wide range of interventional therapies. Neuro procedures – and particularly stroke treatment – benefit from artifact reduction in the basal part of the brain with syngo DynaCT Sine Spin and with syngo DynaCT Multiphase. Clinicians can now visualize collateral vessels at 10 different time points. In addition, Lateral Plane Switch enables different specialties, like interventional cardiologists and radiologists, to perform procedures without compromise. The ARTIS icono also introduces a new imaging chain called OPTIQ which allows interventionalist to choose their own image quality preference at a low dose-- regardless of a patient’s BMI. ARTIS icono floor The ARTIS icono floor is a flexible, multi-axis system that functions like a ceiling system. Its easy-to-use, adaptable design supports improved utilization and growth for all types of procedures and patients. Its revolutionary imaging chain, called OPTIQ, allows interventionalist to choose their own image quality preference, for their procedures, at a low dose-- regardless of a patient’s BMI. The system also provides full patient length coverage of 6.8 ft and 6.2 ft lateral coverage for optimal radial access. And with a minimum room size requirement of only 269 sq. ft, the ARTIS icono floor supports budgets and space constraints.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Tube won't fire, heat-capacity alarm.
Root cause: Filament wear, cooling failure.
Symptoms: Image noise, artifacts, low SNR.
Root cause: Flat-panel / sensor aging.
Symptoms: Tube temp alarm, duty-cycle limited.
Root cause: Pump/radiator or coolant loss.
Symptoms: Exposure fails, kV/mA out of range error.
Root cause: HV tank or generator board fault.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Major R&R; track mAs history.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| High-voltage generator | Major service part | $$$$ | Expert-level |
| Workstation RAID / GPU | Periodic replacement | $$$ | Difficult |
| X-ray tube insert Major R&R; track mAs history. | Major service part | $$$$ | Expert-level |
| Flat-panel detector | Major service part | $$$$ | Expert-level |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.