SenTec · Model SDM
Flip feet serving as carrying handle or to adjust angle for improved table-top viewing Mountable on roll/infusion stands, wall mounts/railings, transport incubators, etc.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Transcutaneous CO2 / O2 values off vs a reference, long stabilisation, drift over hours.
Root cause: Old sensor membrane / electrolyte, a sensor past its remembered life, or a calibration / gas-bottle fault.
Symptoms: The monitor prompts for a membrane change or sensor service, or the sensor will not calibrate.
Root cause: Membrane / electrolyte at end of life (typically every ~28-42 days), or a worn sensor.
Symptoms: Calibration fails in the docking station, gas-bottle empty or expired.
Root cause: Empty / expired SenTec Gas bottle, or a docking-station / regulator fault.
Symptoms: No SpO2 / pulse or poor signal from the combined sensor.
Root cause: Sensor optics or cable fault.
Symptoms: Short battery run time; the sensor will not stay attached (ear clip / adhesive ring).
Root cause: Aged battery, or worn attachment accessories.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Change per the schedule (e.g. every 28-42 days).
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Sensor membrane-change kit & electrolyte Change per the schedule (e.g. every 28-42 days). | Consumable | $$ | Fairly easy |
| tcpO2/tcpCO2 sensor | Frequent wear item | $$$$ | Easy |
| SenTec Gas bottle | Frequent wear item | $$ | Easy |
| Ear clip / adhesive attachment rings | Consumable | $ | Easy |
| Battery | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Device requires the use of an Isolation Transformer in combination with the SDM in home care settings to comply with applicable safety standards.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio