Physio-Control · Model Lifepak 20e
The 20e is highly intuitive to use, making it easy for infrequent AED trained responders to quickly understand and use The closed door reduces the "confusion factor" and ensures basic responders are presented with only the controls they require to facilitate fast and easy operation The proven Shock Advisory System™ from Physio-Control guides users through 1-2-3 step operation, with loud voice prompts and clear, simple graphics cprMAX™ technology provides a resuscitation platform that can increase CPR hands-on time and minimize delays between CPR and shock Clinically advanced and packed with power When the code team arrives, the 20e easily converts to manual mode – with the easy push of a latch the 20e automatically converts to a manual defibrillator For quick and effective clinical decisions, more advanced monitoring parameters such as ECG, external pacing and pulse oximetry are displayed clearly through color-matched waveforms and values MASIMO SET® pulse oximetry offers accurate and stable oxygen saturation monitoring, for quick and effective clinical decisions under conditions of both active movement and low perfusion Lithium-ion battery technology provides extended operating time, making it's easy to transport patients from one area of the hospital to another.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Delivered energy outside tolerance on the analyser, will not charge, or charges slowly.
Root cause: HV capacitor / charging-circuit ageing, or an output-relay fault.
Symptoms: Short standby, readiness indicator shows service, unit fails the self-test on battery.
Root cause: Aged battery pack or a charging fault.
Symptoms: "Check pads", "connect electrodes", no ECG through pads, poor paceing capture.
Root cause: Worn therapy cable, expired pads, or a paddle / connector fault.
Symptoms: A monitoring parameter drops or is implausible.
Root cause: Patient cable / sensor or the parameter module.
Symptoms: No printout, event data will not transfer.
Root cause: Thermal recorder or the data / Wi-Fi interface.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Battery pack | Periodic replacement | $$ | Easy |
| Therapy / defib cable | Periodic replacement | $$ | Fairly easy |
| Multifunction pads | Frequent wear item | $ | Easy |
| HV / charging board | Major service part | $$$$ | Difficult |
| Thermal recorder / print head | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.