XX XX SelfTest ErrAn error occurred to the XX module, or there is a problem with the communications between the module and the host. Restart the equipment.
Restart the equipment.
XX OverrangeThe measured XX value is not within the specified range for XX measurement. Contact our Customer Service Department.
Contact our Customer Service Department.
ECG Lead OffThe ECG electrode has become detached from the patient or the lead wire has become disconnected from the trunk cable. Check the connection of the electrodes and leadwires.
Check the connection of the electrodes and leadwires.
ECG YY Lead OffYY represents the leadwires V, LL, LA, and RA, as per AHA standard, or C, F, L and R as per IEC standard.
Pads/Paddles offThe pads/paddles have been detached from the patient or the therapy cable is loose. Check that the pads/paddles and therapy cable are properly connected.
Check that the pads/paddles and therapy cable are properly connected.
ECG NoiseThe ECG signal is noisy. Check for any possible sources of signal noise form the area around the cable and electrode, and check the patient for excessive motion.
ECG Signal InvalidECG amplitude is so low that ECG signal is undetectable. Check for any possible source of interference from the area around the cable and electrode; check the patient’s condition.
SpO2 Sensor OffThe SpO2 sensor has become detached from the patient or the module, or there is a fault with the Sp O2 sensor, or an unspecified Sp O2 sensor has been used. Check the sensor application site and the sensor type, and make sure the sensor is not damaged. Reconnect the sensor or use a new sensor.
Reconnect the sensor or use a new sensor.
SpO2 Too Much LightThere is too much light on the Sp O2 sensor. Move the sensor to a place with lower level of ambient light or cover the sensor to minimize the ambient light.
Move the sensor to a place with lower level of ambient light or cover the sensor to minimize the ambient light.
SpO2 Low SignalThe Sp O2 signal is too low or too weak. Check the patient’s condition and change the sensor application site. If the error persists, replace the sensor.
Check the patient’s condition and change the sensor application site. If the error persists, replace the sensor.
SpO2 Weak Signalthe patient’s condition and change the sensor application site. If the error persists, replace the sensor.
If the error persists, replace the sensor.
SpO2 InterferenceThe Sp O2 signal has been interfered. Check for any possible sources of signal noise form the area around the sensor, and check the patient for excessive motion.
SpO2 Board FaultThere is a problem with the Sp O2 measurement board. Do not use the module and contact our Customer Service Department.
Do not use the module and contact our Customer Service Department.
NIBP Loose CuffThe NIBP cuff is not properly connected, or there is a leak in the airway.
Check the NIBP cuff and pump for leakages.
NIBP Pneumatic LeakCheck the NIBP cuff and pump for leakages.
Check the NIBP cuff and pump for leakages.
NIBP Cuff Type WrongThe cuff type applied mismatches the patient category. Verify the patient category and replace the cuff.
Verify the patient category and replace the cuff.
NIBP Air Press. ErrAn error occurred to the air pressure. Verify that the equipment application site meets the environmental requirements and check if there is any source that affects the air pressure.
Verify that the equipment application site meets the environmental requirements and check if there is any source that affects the air pressure.
NIBP Weak SignalThe patient’s pulse is weak or the cuff is loose. Check the patient’s condition and change the cuff application site. If the problem persists, change the cuff.
Check the patient’s condition and change the cuff application site. If the problem persists, change the cuff.
CO2 Sensor High TempCheck, stop using or replace the sensor.
Check, stop using or replace the sensor.
CO2 OcclusionThe airway or watertrap was occluded. Check the airway and remove the occlusion.
Check the airway and remove the occlusion.
CO2: Change WatertrapChange the watertrap.
Change the watertrap.
CO2 Watertrap MismatchCheck the patient category, replace a matched watertrap.
Check the patient category, replace a matched watertrap.
CO2 No WatertrapCheck the watertrap connections.
Check the watertrap connections.
CO2 Zero FailedCheck the CO2 connections. After the sensor’s temperature becomes stabilized, perform a zero calibration again.
Check the CO2 connections. After the sensor’s temperature becomes stabilized, perform a zero calibration again.
CO2 Module ErrorThere is a problem with the CO2 module, or a problem with the communications between the host and the CO2 module. Restart the equipment.
Restart the equipment.
CPR Sensor ErrThere is a self-test error or communication problem with the CPR sensor. Contact our Customer Service Department.
Contact our Customer Service Department.
CPR Sensor Low BatteryThe battery power of the CPR sensor is low. Charge the battery by connect the CPR sensor to the equipment.
Charge the battery by connect the CPR sensor to the equipment.
CPR Sensor Need ServiceThe compressions using the CPR sensor exceed the expected numbers. Contact our Customer Service Department.
Contact our Customer Service Department.
CPR Sensor Cable FaultAn error occurred to the CPR sensor cable. Replace the CPR sensor cable.
Replace the CPR sensor cable.
Change CPR Sensor BatteryThe CPR sensor battery is aging. Contact our Customer Service Department.
Contact our Customer Service Department.
CPR Sensor Bat. Charge ErrThe CPR sensor cannot be charged. Contact our Customer Service Department.
Contact our Customer Service Department.
No SpeakerMake sure that the speaker is connected.
Make sure that the speaker is connected.
Power Board Comm ErrAn error occurred to the power board, or there is a problem with the communications between the power board and the host. Restart the equipment.
Restart the equipment.
Keyboard Comm ErrAn error occurred to the keypad board, or there is a problem with the communications between the keypad board and the host. Restart the equipment.
Restart the equipment.
Therapy Module Comm ErrAn error occurred to the therapy module, or there is a problem with the communications between the therapy module and the host. Restart the equipment. If the problem persists, contact our Customer Service Department.
Restart the equipment. If the problem persists, contact our Customer Service Department.
Main Control Selftest ErrThe main control voltage is abnormal. Replace the main control board.
Replace the main control board.
Wifi Module FaultContact our Customer Service Department.
Contact our Customer Service Department.
RT Clock Need ResetReset system time.
Reset system time.
RT Clock ErrAn error occurred to the RTC chip, or the button cell is depleted. Replace corresponding part.
Replace corresponding part.
Memory ErrThere is a problem with the data card. Format the CF card. If the problem persists, contact our Customer Service Department.
Format the CF card. If the problem persists, contact our Customer Service Department.
Last User Test FailedRun a successful user test.
Run a successful user test.
Last Auto Test FailedRun a successful user test again.
Run a successful user test again.
No CMSThe equipment is disconnected from the CMS. Check the network connection.
Check the network connection.