Error Codes
Extracted directly from this unit's own service manual.
ALARM SILENCE SHORTAlarm Silence/Reset key is stuck. Replace the front panel. BLUETOOTH PHONE UNAVAILABLE The X Series unit was unable to establish a Wide Area Network connection. None of the Bluetooth phones paired with the device were detected. To correct the problem, verify the following: The Bluetooth device is paired with the X Series unit. The phone is on and near enough to the device to permit communication. The phone is configured to allow data connection requests (if the phone has such a configuration setting). CELLULAR CONNECT FAILURE The X Series unit was unable to establish a Wide Area Netwo
CO2 CAL FAILURECalibration failed. Retry calibration or continue with old calibration. Replace the CO2 module.
CO2 CAL REMINDERCalibration reminder. Calibrate the CO 2 module soon.
CO2 CAL REQUIREDCalibration due. Calibrate the CO 2 module.
CO2 CHECK FLOWExhaust port blocked. Check for exhaust port occlusion. If problem persists, replace the CO 2 module.
CO2 COM ERROR CODECommunication failure. Check communication cables. If problem persists, replace the CO 2 module.
CO2 DEVICE FAILUREMalfunction or device failure. Replace the CO 2 module. Error Message Explanation Technical Action
CO2 NO FILTERLINENo filterline is detected. If FilterLine is connected and problem persists, replace FilterLine. If problem still persists, replace the CO 2 module.
DEFIB CAP LEAKAGECapacitor bank cannot hold a charge. Replace the capacitor bank module.
DEFIB CHARGE ERROROver voltage was detected. If problem persists, replace the defibrillator board module. DEFIB CHARGE TIMEOUT FAIL The battery did not produce enough current to charge capacitor. Charge the battery. If voltages are okay (from battery and aux), replace the defibrillator board module.
DEFIB DEVICE FAILUREOne of the following issues occurred: Device could not charge up the self-test energy. Discharge circuit failed to discharge voltage within specified time duration. Device failed to deliver correct self-test energy. Check communication cables. If problem persists, replace defibrillator board module.
DEFIB DISCHARGE TEMPDevice is in a hot environment or the discharge circuit is always on. Move device to cooler area. If problem persists, replace the defib board module. DEFIB FATAL CHARGE ERROR Charging error. Replace the defibrillator board module. Error Message Explanation Technical Action X Series Error Messages 9650-001356-01 Rev. D X Series Service Manual 53
DEFIB FIRE FAILUREHigh impedance or low impedance (short). Replace pads. If problem persists, check internal connections. If problem still persists, replace the defibrillator board module.
DEFIB LOW BATTERYBattery is too low to charge defibrillator. Charge the battery. Check battery contacts and clean if necessary. DEFIB NO ENERGY DELIVERED High impedance. No defib energy was delivered. Replace pads. If problem persists, check internal connections. If problem still persists, replace the defibrillator board module. DEFIB PACER DEVICE FAILURE One of the following issues occurred: Error was detected on the defib/pacer board. Interface self-test communication failed. Pacer interface self-test failed. Communication with pacer hardware failed after startup. Check communication cables. If p
DEFIB PHASE FAILUREPhase I or phase II waveform dropout. Replace the defibrillator board module. DEFIB USING DEFAULT CALIBRATION Calibration data lost. Replace the defibrillator board module.
ECG DEVICE FAILUREECG software detected a problem with the ECG data that is most likely a hardware issue. Replace the main board. FP DEFIB CHARGE SHORT Keys are stuck. Replace the front panel. FP DEFIB ENERGY DOWN SHORT Defib energy Select Down (Decrease) key is stuck. Replace the front panel. FP DEFIB ENERGY UP SHORT Defib energy Select Up (Increase) key is stuck. Replace the front panel.
FP DEFIB FIRE SHORTA key or a front panel button is stuck. Replace the front panel. Error Message Explanation Technical Action
HW SHUTDOWNVoltage threshold exceeded the minimum working threshold of the device. There is no warning. Charge the battery.
IBP DEVICE FAILUREIBP software detected a problem with the IBP data that is most likely a hardware issue. Replace the main board. IBP PROBE1/PROBE2/ PROBE3 SHORTED An A/D value was detected that indicated the probe was shorted. Replace the problematic IBP probe. If the problem persists, replace the right side panel. IBP PROBE1/PROBE2/ PROBE3 CIRCUIT FAILURE An unsupported probe was inserted or there is a circuit problem. Unplug the problematic probe, power cycle. If problem persists, replace main board.
NAMESERVICE FAILUREThe reports server hostname could not be resolved via DNS. To correct the problem, verify the following: The server DNS name is entered correctly and completely in the reports configuration. The DNS server(s) are correctly configured (if using a Wi-Fi profile where static IP information was configured). The DHCP server on the Wi-Fi network provides valid and correct DNS information in its response (if using a Wi-Fi profile with DHCP). If using any form of cellular connectivity, retry the operation.
NAV PAD DOWN SHORTDown/Counterclockwise Navigation key is stuck. Replace the front panel.
NAV PAD SELECT SHORTSelect key is stuck. Replace the front panel.
NAV PAD UP SHORTUp/Clockwise Navigation key is stuck. Replace the front panel. Error Message Explanation Technical Action X Series Error Messages 9650-001356-01 Rev. D X Series Service Manual 55
NIBP AIR LEAKMajor air leak or could not reach target pressure. Attempt to identify problems with the hose and/or cuff by doing the following: Check the hose connection to the device. Make sure that the hose is tightly screwed onto the NIBP connector. Ensure that the cuff is appropriately wrapped and is not too loose. Inspect the cuff and hose for leaks. Inspect the O-ring on the NIBP hose connector. Consider replacing the cuff and/or hose. If the problem persists, perform the following procedure: “NIBP Leak Test” on page 40: Inspect internal NIBP tubing for leaks.
NIBP ARTIFACTUnable to determine blood pressure. Check connections; restrict patient movement. Attempt to identify and eliminate the source of artifact. If problem persists, replace the NIBP module. NIBP COMMUNICATION FAILURE There is a communication problem with the NIBP module. Ensure that the J32 connection is secure. If error persists, replace the NIBP module.
NIBP EQUIPMENT CHECKDevice is unable to obtain a reading. Check the hose and cuff connections. If problem persists, perform one or more of the following procedures in Chapter 1: NIBP Leak Test NIBP Deflation Test NIBP Inflation Test
NIBP KINKED HOSEThe cuff inflation was too rapid. There may be kinked tubing or another air obstruction. Check the external NIBP hose for kinks. If problem persists, check the routing of the NIBP tubing inside the device. Error Message Explanation Technical Action
NIBP LOW VOLTAGEInsufficient voltage is supplied to the NIBP module. Check the module voltage in the NIBP Test screen. Inspect the unit or water ingress. If evidence of water ingress is found, allow the water to evaporate while periodically checking the module voltage. If problem persists, replace the NIBP module. NIBP MAX PRESSURE EXCEEDED Pressure limit exceeded. If problem persists, do the following: Check the calibration via the “NIBP Calibration Check” on page 37. Check the routing of the NIBP tubing inside the device for kinks or obstructions. If the NIBP module is out of calibration, replace the
NIBP WEAK PULSEToo few pulses are detected during the reading attempt. Tighten the cuff and ensure appropriate placement. Error Message Explanation Technical Action X Series Error Messages 9650-001356-01 Rev. D X Series Service Manual 57
PACER CURRENT FAULTLack of pacer current. Replace the defibrillator board module.
PACER DEVICE FAILUREGeneration of pacer pulse failed. Replace the defibrillator board module.
PRINTER OFFLINEPrinter module is off-line. If it remains off line, try checking the connectors or power to the printer. If problem persists, replace the printer module.
PRINTER OUT OF PAPERIf there is paper in the device, probably a paper sensor problem. Replace the printer module.
PRINTER PLATEN OPENIf printer door is closed, probably a problem with the micro switch. Replace the printer module. PS DEFIB CHARGE SHORT One of the paddle set keys are stuck. Replace the paddle set or MFC.
PS DEFIB ENERGY DOWNPaddle set keys are stuck. Replace the paddle set or MFC. PS DEFIB ENERGY DOWN SHORT Paddle set energy down select button is stuck. Replace the paddles.
PS DEFIB ENERGY UPPaddle set keys are stuck. Replace the paddle set or MFC.
PS DEFIB FIRE SHORTPaddle set keys are stuck. Replace the paddle set or MFC. PS DEFIB RECORDER SHORT One of the paddle set keys are stuck. Replace the paddle set or MFC. QUICK ACCESS SW 1(to SW 7) SHORT One of the quick access keys (key 1 to key 7) is stuck. Replace the front panel.
SCREEN CYCLE SHORTDisplay/home key is stuck. Replace the front panel. Error Message Explanation Technical Action
SERVER UNREACHABLEAlthough a WAN connection was established, the reports server connection could not be established or the connection was lost during a transfer. Retry the operation. SERVICE ERROR <CODE> While communicating with the remote server, there was an error. Retry the operation.
SNAPSHOT SHORTSnapshot key is stuck. Replace the front panel.
SPO2 COM ERROR CODECommunication error. Check communication cables. If problem persists, replace the SpO2 module.
SPO2 DEVICE FAILURECommunication failure or device malfunction. Check the SpO2 communication cables. If problem persists, replace the SpO 2 module.
SPO2 SENSOR FAILUREDefective sensor or sensor error. Replace the SpO 2 sensor. SPO2 UNKNOWN RECOVERY NUM Recovery error. Replace the SpO 2 sensor. SW SHUTDOWN WARNING Indication that the battery has reached the software shutdown capacity. Charge the battery. TEMP1 (or TEMP2) CIRCUIT FAILURE The software detected an out-of-range A/D converter. Remove probe then power cycle. If problem persists, replace the main board. Error Message Explanation Technical Action X Series Error Messages 9650-001356-01 Rev. D X Series Service Manual 59
TEMP DEVICE FAILURETemp software detected a problem with the temperature data that is most likely a hardware issue. Replace the main board. TEMP PROBE1 (or PROBE2) SHORTED An A/D value was detected that indicated the probe was shorted. Replace the temperature probe or adapter cable. If problem persists, replace the right side panel.
TRANSMISSION FAILEDThe communication operation failed. This message displays until a new operation is started. To correct the problem, verify the following: Wireless communication is enabled on the X Series unit. The Wi-Fi settings are correct in the Communications setup menu. The ZOLL server is configured correctly. Your cellular device is configured correctly. The X Series unit is within range of the wireless server. WAN CONNECTION UNAVAILABLE The X Series unit was unable to establish a WAN connection. To correct the problem, verify the following: Wireless communication is enabled on the X Se