Philips Healthcare · Model M3046A M3
Patient demographics, measurement data and trends, acquired and stored in the MMS, can be transferred to the M-Series (M3/M4) and Philips Intellivue (MP60/70/90) patient monitors. EASI derived 12 lead ECG using 5 electrode cable set. Multi lead arrhythmia. Up to 3 channels of ECG with basic arrhythmia and ST analysis capability. FAST-SpO2 measurement, using Fourier Artifact Suppression Technology, for accurate performance even with low perfusion. Respiration, non-invasive and invasive blood pressure and temperature measurements. Mainstream or sidestream CO2 a second invasive blood pressure and temperature measurements and enhanced arrhythmia capability, available optionally. Choice of serial interface for a dedicated external strip chart recorder (as default), or IrDA interface for connecting tot a local printer. 3-wave or 4-wave displays are available. Trend data, with manual and automatic event storage, together with a range of report styles help you track and document your patients progress. Color display with a wide viewing angle. Optional battery informs you of remaining capacity, and allows longer transport times. It can be easily exchanged during transport with no loss of settings. Touch bar human interface - quick to learn and easy to use. Rubber bezel protector, available in four colors, gives additional protection in transport scenarios.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: ECG, IBP, SpO2, temperature, CO2 or CO intermittently drops or is implausible.
Root cause: Worn patient cable / sensor, or a parameter-module / input-board fault.
Symptoms: NIBP times out, cuff will not inflate / hold, readings implausible.
Root cause: Internal NIBP pump or valve, a cuff / hose leak, or a pressure-transducer fault.
Symptoms: The host does not display a plugged-in module or a built-in parameter.
Root cause: Docking / connector fault or an internal board fault.
Symptoms: Blank, dim or distorted display; unresponsive touchscreen.
Root cause: Display, backlight, or controller fault.
Symptoms: Short battery run time, will not hold charge, will not power on.
Root cause: Aged battery pack or a power-supply fault.
Symptoms: Monitor not shown at the central station, admit fails, intermittent drop.
Root cause: Wired / Wi-Fi network configuration or a communication board.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable set (ECG / IBP / SpO2 / temp) | Frequent wear item | $$ | Easy |
| NIBP cuff and hose | Frequent wear item | $ | Easy |
| Internal NIBP pump / valve | Major service part | $$$ | Difficult |
| Parameter module / input board | Major service part | $$$ | Difficult |
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Display / touchscreen assembly | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio