Philips Healthcare · Model IntelliVue MP80
Networked advanced care Dynamic Wave area features waves that automatically adjust in size depending on the number of waves configured. Capture, review, and store diagnostic 12-lead ECGs at the monitor before sending them to the IntelliVue Information Center. You can also locally print in a harmonized layout. Each NBP measurement now generates a column in the vital signs trend table. Measurements for other values are added to provide a comprehensive vital signs data set for the NBP measurement time, offering a more complete picture. The Smart Alarm Delay algorithm helps reduce the number of pulse oximetry nuisance alarms, allowing you to focus your attention where needed. Customized viewing options let you see and analyze data in graphical or numerical formats, and organize every on-screen element from waveforms to data labels as desired. Choice of display including medical-grade XGA and SXGA stand-alone, off-the-shelf models or Philips 38cm (15?)/43cm (17?) color LCD XGA display with touchscreen operation. Touchscreen navigation is available with compatible displays. Many functions are accessible through simple one-touch commands. High performance CPU Individual clinical measurement modules expand Philips clinical measurement offering. IntelliVue MP80 works with one Flexible Module Server plus the Multi-Measurement Server and extensions to increase measurement capabilities to match patient acuity. Remote SpeedPoint for easy data entry and navigation. Can be used together with a remote alarm device for visible and audible alerts.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Flat or noisy ECG, frequent lead-off alarms.
Root cause: Worn electrodes, loose leads, or dry gel.
Symptoms: No pleth, SpO2 drops to --- or false low.
Root cause: LED/photodiode aging or cable damage.
Symptoms: Cuff fails to inflate, erroneous BP, long cycle.
Root cause: Pump/valve wear or cuff leak.
Symptoms: Dim, flickering, or unresponsive screen.
Root cause: Backlight or digitizer failure.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
High-turnover; keep inventory.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Rechargeable backup battery | Periodic replacement | $$ | Fairly easy |
| Display / touch assembly | Periodic replacement | $$$ | Difficult |
| ECG patient cable & lead set High-turnover; keep inventory. | Frequent wear item | $$ | Easy |
| SpO2 sensor (adult/ped/neo) | Frequent wear item | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →