Philips Healthcare · Model IntelliVue MMS X2
• One of the lightest, smallest and rugged critical care transport monitors available • An MMS with display, alarm capability, 3-hour removable battery, and extended trends • Crisp, colorful touchscreen with 3.5” display with seamless electronic recording • Connects to the IntelliVue XDS Large Display to add a high-resolution larger bedside display to see clearly see vital information at a glance and to provide remote-control functionality • Transmits data wired or wirelessly to the IntelliVue Information Center, or just plug and play to upload data to another IntelliVue monitor • Wireless telemetry as a parameter (WTAAP) expands monitoring capabilities to provide untethered in-room access to ECG/SpO2 data in near-real time • Philips Green Flagship product, meeting rigorous environmental standards in manufacturing and use
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Flat or noisy ECG, frequent lead-off alarms.
Root cause: Worn electrodes, loose leads, or dry gel.
Symptoms: Cuff fails to inflate, erroneous BP, long cycle.
Root cause: Pump/valve wear or cuff leak.
Symptoms: Short backup time, battery alarm on transport.
Root cause: Aged backup battery.
Symptoms: Dim, flickering, or unresponsive screen.
Root cause: Backlight or digitizer failure.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
High-turnover; keep inventory.
Consumable per patient.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| ECG patient cable & lead set High-turnover; keep inventory. | Frequent wear item | $$ | Easy |
| SpO2 sensor (adult/ped/neo) | Frequent wear item | $$ | Easy |
| NIBP cuff & hose Consumable per patient. | Periodic replacement | $ | Easy |
| Rechargeable backup battery | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.