Penlon · Model Prima SP2
Advanced easy to use features for maximum patient safety Comprehensive choice of layout through modular open architecture Ultra low flow anesthesia Advanced AV-S Ventilator with six ventilation modes A200SP Absorber with ventilator interface Life-Care’ or optional ‘Life-Care Plus’ Warranty and Customer Care scheme Seamless workstation integration Two vaporizer station as standard Up to four gases Six ventilation modes Large color touchscreen.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Breathing-system / absorber leak test fails, machine not ready for use.
Root cause: Breathing-system / absorber leak, a worn APL valve or bag arm, a CO2-absorber canister seal, or a vaporiser back-bar leak.
Symptoms: O2 flow not matching the setting, flowmeter float sticking, hypoxic-guard / O2-N2O ratio fault, no O2 flush.
Root cause: Worn flowmeter (rotameter), a hypoxic-guard link fault, or a flush-valve fault.
Symptoms: Agent output does not match the dial, back-bar leak, dial stiff.
Root cause: Vaporiser seating / O-ring, or an internal vaporiser fault.
Symptoms: Set tidal volume not delivered, bellows will not fill, PEEP not held.
Root cause: Bellows leak, drive-gas valve, or a flow-sensor fault.
Symptoms: O2-failure whistle / alarm, anti-hypoxia device cuts N2O.
Root cause: Loss of O2 pipeline / cylinder pressure, or a pressure-sensor fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| APL valve / breathing-system gaskets | Periodic replacement | $$ | Fairly easy |
| CO2 absorbent / canister seals | Consumable | $ | Easy |
| O2 flowmeter (rotameter) | Major service part | $$ | Moderate |
| Vaporiser O-ring / back-bar seal | Periodic replacement | $ | Fairly easy |
| Ventilator bellows | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.