GE HealthCare · Model Aestiva/5
Superior ventilation Volume mode, pressure mode, electronic PEEP Tidal volume compensation One motion from mechanical to manual mode Two key presses to total standby: end case Cardiac bypass case mode Open systems architecture Lower overall height User configurable drawers/shelving Innovative patient breathing system Eight machine hoses/cables integrated “No tools” disassembly of components Autoclavable and latex-free Responsive location of common gas outlet Improved low flow/reduced life cycle costs Fresh gas flow compensation - automatically Smooth, faster acting fresh gas flow control Minimum O2 flow of 50 mL Dual air flow tube for low flow
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Delivered agent concentration off, low/high agent alarm.
Root cause: Worn bellows or O-ring; agent contamination.
Symptoms: O2 low-pressure alarm, inability to ventilate.
Root cause: Pipeline disconnect or failed O2 flush valve.
Symptoms: Low minute volume, volume loss, leak alarm.
Root cause: Loose connections, torn bellows, cracked hose.
Symptoms: Rising EtCO2, patient rebreathing.
Root cause: CO2 absorbent saturated.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable - replace on schedule.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Breathing circuit / bellows | Periodic replacement | $$ | Fairly easy |
| CO2 absorber canister Consumable - replace on schedule. | Consumable | $ | Easy |
| Scavenger interface & hose | Periodic replacement | $$ | Fairly easy |
| O2 pressure regulator | Periodic replacement | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.