Partec · Model CyFlow ML
Software Partec FloMax® FCM Data Acquisition and Analysis Software Windows™ based Partec FloMax® software is including multiple online gating, logical gates, and colour gating. N-colour online and offline software compensation is an other unique feature of FloMax® software. Data may be acquired with compensation or as uncompensated raw data. Either way compensation can be done, undone or changed after analysis on stored data. For data elaboration quadrant analysis, cell cycle analysis, peak analysis and other sophisticating features are available. An optional report function allows the generation and the individual design of analysis report sheets as Word or Excel documents. FloMax® data files are stored in FCS standard and therefore fully compatible to other commercial FCM data elaboration software solutions. FloMax® software already is pre-installed on the Windows™ PC delivered as part of the CyFlow® ML system. The Partec CyFlow® ML is supplied with additional 3 offline laboratory licenses free of charge, offering a cost-effective solutions also for working groups with need for offline FCM analysis software installations. Options Flexible and Modular Options and Upgrades for Partec CyFlow® ML Up to five light sources (e. g. laser light sources, modern high power LEDs and UV mercury arc lamp) and up to sixteen optical detectors yield an unlimited matrix of configurations for the Partec CyFlow® ML. Please contact the Partec Science and Application team for your specific requirements.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Bead CVs out of spec, scatter or fluorescence signal low or noisy, compensation drift.
Root cause: Flow-cell contamination or an air bubble, laser degradation / misalignment, or an optics / PMT fault.
Symptoms: Unstable event rate, boost / prime failures, sheath or waste alarm, clogs.
Root cause: Sheath / waste tank issue, a peristaltic / vacuum fault, or a partially blocked sample line / flow cell.
Symptoms: No or low event rate, probe-position error, carryover.
Root cause: Clogged or bent sample probe, or a clot.
Symptoms: A laser-line-specific parameter fails, laser-error message.
Root cause: Laser at end of life or a power-supply / interlock fault.
Symptoms: Acquisition hangs, data will not export, instrument not recognised.
Root cause: Workstation PC or the acquisition software / link.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Flow cell | Periodic replacement | $$$ | Difficult |
| Sample probe / SIP | Periodic replacement | $$ | Moderate |
| Peristaltic pump tubing | Frequent wear item | $ | Moderate |
| Sheath / waste filters | Frequent wear item | $ | Easy |
| Laser | Periodic replacement | $$$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.