Nihon Kohden · Model BSM-9101
• MULTI connectors • 15-inch high resolution color TFT LCD display. A local purchased display can also be used. • Dual display with independant display mode • Customizable screen layouts for easiest viewing at different sites • Flexible control of the monitor by touch screen, remote controller, mouse, and keyboard • Large data storage capacity and powerful review for waveforms, alarm history, arrhythmia and ST files, and trend data • 12-lead ECG analysis • Gas measurement with AG-920R Multigas Unit • Different parameters are available by adding different modules • Connection to external equipment with an optional interface • Networking - Built-in LAN port allows Ethernet networking with an LS-NET patient monitoring network to send waveforms and numerical data to the central monitor • Revolutionary PWTT technology - This monitor incorporates a new parameter, PWTT (pulse wave transit time), which can have correlation with blood pressure. The monitor continuously calculates PWTT from ECG and SpO2. During periodic NIBP measurement, if PWTT exceeds the threshold it triggers NIBP measurement. This increases the chance to detect sudden change in blood pressure between periodic NIBP measurements.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: No pleth, SpO2 drops to --- or false low.
Root cause: LED/photodiode aging or cable damage.
Symptoms: Cuff fails to inflate, erroneous BP, long cycle.
Root cause: Pump/valve wear or cuff leak.
Symptoms: Short backup time, battery alarm on transport.
Root cause: Aged backup battery.
Symptoms: Dim, flickering, or unresponsive screen.
Root cause: Backlight or digitizer failure.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
High-turnover; keep inventory.
Consumable per patient.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Display / touch assembly | Periodic replacement | $$$ | Difficult |
| ECG patient cable & lead set High-turnover; keep inventory. | Frequent wear item | $$ | Easy |
| SpO2 sensor (adult/ped/neo) | Frequent wear item | $$ | Easy |
| NIBP cuff & hose Consumable per patient. | Periodic replacement | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.