Nihon Kohden · Model AG-920RA
Automatic identification of five anesthetic agents Simultaneous display of inspired and expired concentrations of 2 anesthetic agents Displays inspired and expired O2, CO2, and N2O 45 second warm-up Automatic adult and neonatal sampling rates Automatic self zero calibration All controls and information display is provided on one monitor with user-friendly touchscreen operation Quick access anesthesia windows provide the following information: [5 waveforms (O2, CO2, N2O, and 2 agents) with large, easy-to-view numerical data color coded to match each waveform Comprehensive graphical trends of all parameters Tabular data list shows minute-by-minute values of all anesthetic agents, gases and vital signs User-selectable alarms for all parameters.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: CO2 / agent / N2O / O2 values biased, will not calibrate, zero drift.
Root cause: Sample-cell / bench contamination, an aged O2 cell, or a calibration-gas / calibration issue.
Symptoms: "Occlusion" or "sample line blocked", low sample flow, no waveform.
Root cause: Blocked sampling line or a saturated / full water trap (Nafion / moisture trap).
Symptoms: Low or no sample flow with a clear line, pump noisy or stopped.
Root cause: Worn sample pump.
Symptoms: O2 reads wrong or will not calibrate.
Root cause: Aged galvanic O2 cell or a paramagnetic-sensor fault.
Symptoms: Host monitor does not see the gas module, intermittent dropout.
Root cause: Module connector / docking or an internal board fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Water trap / moisture trap | Consumable | $ | Easy |
| Sampling line | Consumable | $ | Easy |
| Sample pump | Periodic replacement | $$ | Moderate |
| O2 sensor / cell | Frequent wear item | $$ | Fairly easy |
| Gas bench / sample cell | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.