Encapsulated Seat Technology (EST) ensures superior performance, accuracy and resistance to ‘surging’ Large diameter diaphragm gives a performance that is normally only found in double stage regulators Triple filter system increases the regulator performance by preventing foreign particles from getting in contact with the high pressure gas or the regulator seat No tools are required to routinely disconnect the regulator from the cylinder valve A safety relief valve is built in to the regulator to protect staff and patients in the event of high pressure failures Tamper resistant design to ensure only qualified personnel can access the regulator Color coded pressure gauges and identification labels for quick recognition of the gas used by the regulator Gas specific inlet fittings and outlet fittings to prevent fitting to the wrong gas type.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: CO2 / agent / N2O / O2 values biased, will not calibrate, zero drift.
Root cause: Sample-cell / bench contamination, an aged O2 cell, or a calibration-gas / calibration issue.
Symptoms: "Occlusion" or "sample line blocked", low sample flow, no waveform.
Root cause: Blocked sampling line or a saturated / full water trap (Nafion / moisture trap).
Symptoms: Low or no sample flow with a clear line, pump noisy or stopped.
Root cause: Worn sample pump.
Symptoms: O2 reads wrong or will not calibrate.
Root cause: Aged galvanic O2 cell or a paramagnetic-sensor fault.
Symptoms: Host monitor does not see the gas module, intermittent dropout.
Root cause: Module connector / docking or an internal board fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Water trap / moisture trap | Consumable | $ | Easy |
| Sampling line | Consumable | $ | Easy |
| Sample pump | Periodic replacement | $$ | Moderate |
| O2 sensor / cell | Frequent wear item | $$ | Fairly easy |
| Gas bench / sample cell | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.