Nihon Kohden · Model GF-210R/GF-220R
Measures respiratory gas concentrations Measures respiratory volume with the GF-220R model Designed primarily for use in operating rooms Continuously monitors O‚, CO‚, N‚O, and volatile anesthetic agents, including halothane, enflurane, isoflurane, sevoflurane, and desflurane Continuously monitors airway flow, airway pressure, and airway volume on the GF-220R model Identifies inspiratory and expiratory phases using real-time gas concentration data Calculates respiratory rate and determines gas concentrations during inhalation and exhalation Automatically identifies five anesthetic agents Begins measurement approximately 60 seconds after power-up Uses a low sampling flow rate of 200 mL/min, making it suitable for volatile anesthetic measurement and neonatal monitoring Offers automatic or user-adjustable flow-rate settings Supports both horizontal and vertical mounting
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Delivered agent concentration off, low/high agent alarm.
Root cause: Worn bellows or O-ring; agent contamination.
Symptoms: Low minute volume, volume loss, leak alarm.
Root cause: Loose connections, torn bellows, cracked hose.
Symptoms: High scavenger pressure, agent smell in OR.
Root cause: Obstructed scavenger line or failed regulator.
Symptoms: Rising EtCO2, patient rebreathing.
Root cause: CO2 absorbent saturated.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable - replace on schedule.
Requires periodic calibration.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| CO2 absorber canister Consumable - replace on schedule. | Consumable | $ | Easy |
| Scavenger interface & hose | Periodic replacement | $$ | Fairly easy |
| O2 pressure regulator | Periodic replacement | $$$ | Difficult |
| Vaporizer (agent-specific) Requires periodic calibration. | Major service part | $$$$ | Expert-level |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.