Nihon Kohden · Model GF-210R/GF-220R
Measures respiratory gas concentrations Measures respiratory volume with the GF-220R model Designed primarily for use in operating rooms Continuously monitors O‚, CO‚, N‚O, and volatile anesthetic agents, including halothane, enflurane, isoflurane, sevoflurane, and desflurane Continuously monitors airway flow, airway pressure, and airway volume on the GF-220R model Identifies inspiratory and expiratory phases using real-time gas concentration data Calculates respiratory rate and determines gas concentrations during inhalation and exhalation Automatically identifies five anesthetic agents Begins measurement approximately 60 seconds after power-up Uses a low sampling flow rate of 200 mL/min, making it suitable for volatile anesthetic measurement and neonatal monitoring Offers automatic or user-adjustable flow-rate settings Supports both horizontal and vertical mounting.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: CO2 / agent / N2O / O2 values biased, will not calibrate, zero drift.
Root cause: Sample-cell / bench contamination, an aged O2 cell, or a calibration-gas / calibration issue.
Symptoms: "Occlusion" or "sample line blocked", low sample flow, no waveform.
Root cause: Blocked sampling line or a saturated / full water trap (Nafion / moisture trap).
Symptoms: Low or no sample flow with a clear line, pump noisy or stopped.
Root cause: Worn sample pump.
Symptoms: O2 reads wrong or will not calibrate.
Root cause: Aged galvanic O2 cell or a paramagnetic-sensor fault.
Symptoms: Host monitor does not see the gas module, intermittent dropout.
Root cause: Module connector / docking or an internal board fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Water trap / moisture trap | Consumable | $ | Easy |
| Sampling line | Consumable | $ | Easy |
| Sample pump | Periodic replacement | $$ | Moderate |
| O2 sensor / cell | Frequent wear item | $$ | Fairly easy |
| Gas bench / sample cell | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.