Mortara · Model Surveyor S19
Surveyor S19 operates with a conventional 3 or 5-lead ECG patient cable connection and with the 12-lead ECG patient cable, AM12. high-resolution 18.5" touchscreen display, offers a comprehensive solution for advanced monitoring needs. Externally mounted, thermal printing recorder solution for added flexibility. Capability of interfacing directly with EHR systems to annotate real-time vital signs and waveforms in patient records. Interfaces to the Surveyor Central Monitoring System using wired technology for a complete patient monitoring solution. Covidien's unique Microstream® (CO2) Capnography produces accurate readings in a rapid timeframe so that clinicians can view their patient’s CO2 status faster. Microstream’s 50 mL/min patient sampling and efficient moisture filtering eliminates the need for messy water separation fixtures which often accompany costly liquid collection bottles. Large, clear display supports up to 8 waveforms. Touchscreen operation simplifies workflow by eliminating levels of menus and complexity. Intuitive user interface allows clinician to quickly adjust parameter settings. Ideal for mounting to a wall, rolling stand or counter top, with quick disconnect capability. Occupies minimal space in crowded patient rooms and treatment areas Long battery Run time.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: ECG, IBP, SpO2, temperature, CO2 or CO intermittently drops or is implausible.
Root cause: Worn patient cable / sensor, or a parameter-module / input-board fault.
Symptoms: NIBP times out, cuff will not inflate / hold, readings implausible.
Root cause: Internal NIBP pump or valve, a cuff / hose leak, or a pressure-transducer fault.
Symptoms: The host does not display a plugged-in module or a built-in parameter.
Root cause: Docking / connector fault or an internal board fault.
Symptoms: Blank, dim or distorted display; unresponsive touchscreen.
Root cause: Display, backlight, or controller fault.
Symptoms: Short battery run time, will not hold charge, will not power on.
Root cause: Aged battery pack or a power-supply fault.
Symptoms: Monitor not shown at the central station, admit fails, intermittent drop.
Root cause: Wired / Wi-Fi network configuration or a communication board.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable set (ECG / IBP / SpO2 / temp) | Frequent wear item | $$ | Easy |
| NIBP cuff and hose | Frequent wear item | $ | Easy |
| Internal NIBP pump / valve | Major service part | $$$ | Difficult |
| Parameter module / input board | Major service part | $$$ | Difficult |
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Display / touchscreen assembly | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Martara Instrument, Inc. has recently became aware of a potential safety hazard involving our Surveyor rand S12 S19 Patient Monitors. A hazardous situation may be created when mounting these monitors using the 75mm VESA mounting hole pattern provided in the battery cover of these monitors. If the mounting screws provided by Mortara are not used and if the mounting screws used are longer than the p