Mindray · Model Passport V
The large 12.1” display is highly visible from most vantage points. ECG, NIBP, temperature, respiration and Masimo SET® motion-tolerant SpO2 as its standard configuration 96 hours of patient data is provided in a list or graphic trend format so you can easily review your patient’s history\ Comprehensive calculations packages including both IV drug and hemodynamic data allow for quick and automated calculations. Titration tables and calculations can be printed to the internal recorder or a networked laser printer to satisfy long-term documentation needs. Remote View mode with 8-bed alarm notification enables clinicians to monitor multiple patients without leaving the bedside The innovative quick-release mount makes transport a bit easier. Remove the monitor from its wall mount or rolling stand using the one-touch mechanism and you’re ready for transport. No need to disconnect patient cables. Pair it with the Gas Module 3™ multi-gas analyzer to create a powerful, yet compact O.R. monitoring solution
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Flat or noisy ECG, frequent lead-off alarms.
Root cause: Worn electrodes, loose leads, or dry gel.
Symptoms: No pleth, SpO2 drops to --- or false low.
Root cause: LED/photodiode aging or cable damage.
Symptoms: Cuff fails to inflate, erroneous BP, long cycle.
Root cause: Pump/valve wear or cuff leak.
Symptoms: Dim, flickering, or unresponsive screen.
Root cause: Backlight or digitizer failure.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
High-turnover; keep inventory.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Rechargeable backup battery | Periodic replacement | $$ | Fairly easy |
| Display / touch assembly | Periodic replacement | $$$ | Difficult |
| ECG patient cable & lead set High-turnover; keep inventory. | Frequent wear item | $$ | Easy |
| SpO2 sensor (adult/ped/neo) | Frequent wear item | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →