Mindray · Model Datascope Accutorr plus 2.5
Trend memory up to 100 measurements. High contrast LCD (advanced models only). Universal mounting options. Weighs about 11-pounds.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: No pleth, SpO2 drops to --- or false low.
Root cause: LED/photodiode aging or cable damage.
Symptoms: Cuff fails to inflate, erroneous BP, long cycle.
Root cause: Pump/valve wear or cuff leak.
Symptoms: Short backup time, battery alarm on transport.
Root cause: Aged backup battery.
Symptoms: Dim, flickering, or unresponsive screen.
Root cause: Backlight or digitizer failure.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
High-turnover; keep inventory.
Consumable per patient.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Display / touch assembly | Periodic replacement | $$$ | Difficult |
| ECG patient cable & lead set High-turnover; keep inventory. | Frequent wear item | $$ | Easy |
| SpO2 sensor (adult/ped/neo) | Frequent wear item | $$ | Easy |
| NIBP cuff & hose Consumable per patient. | Periodic replacement | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
NIBP valve assembly board subject to this recall includes enhancements that are not included in the current marketing clearance:1. Anti-motion feature 2. NIBP hose blockage identification function
An issue affecting the caster mount on the AS3000 System have been identified, specifically reports of the caster mount breaking, causing the wheel to fall off and the unit to tip.
SmarTemp Disposable Probe covers may contain microscopic holes in the molded seam that would allow liquids or body fluids inside the cover and potentially contaminate the temperature probe.
Software Anomaly: Software anomaly can occur when the Panorama is attempting to play an audio file, either a physiological or technical alarm sound, at which time the system software may cause the system to freeze due to an interaction in the alarm system.
Two issues have been identified with the AS3000 unit. 1. The threadlocker of the caster (wheels) may not have been utilized, which may allow the caster to loosen and possibly separate from the unit. 2. The use of select brands of pre-pack absorber in the absorber canister of the AS3000 has been associated, in some cases, with gas leakage around the pre-pack, rather than through the absorber mate