Mindray · Model Accutorr 7
Large 8.4″ back-lit high resolution TFT display provides clear view Intuitive touchscreen offers easy access to information Available SpO2 technology includes Masimo SET® (standard) or Nellcor Oximax® PI (Perfusion Index) from SpO2 can guide caregivers to best measurement location Manual, automatic and customizable NIBP measurement modes offer flexibility for various clinical applications Standard MEWS (Modified Early Warning Score) offers effective support for clinical decisions Manual input of vitals such as respiration rate, pain level, glucose, fluids and degree of consciousness Optional Mindray SmarTemp® or Welch Allyn® SureTemp® Plus temperature Low flow sidestream CO2 provides measurements of FiCO2, etCO2, awRR with a CO2 waveform Stores up to 5000 measurements, accessible by patient ID for easy review Optional barcode scanner can speed patient ID input Choice of hardwired or wireless EMR connectivity options Integration with ADT and results messaging easily accomplished via Mindray’s powerful eGateway Direct HL7 interface capabilities Clinician authentication through Mindray MLDAP Server connecting with the hospital Active Directory (AD) Centrally monitor patients on the Accutorr 7 with the Mindray BeneVision Central Station Supports nurse call interface Standard RS-232 interface for third party connectivity solutions.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: ECG, IBP, SpO2, temperature, CO2 or CO intermittently drops or is implausible.
Root cause: Worn patient cable / sensor, or a parameter-module / input-board fault.
Symptoms: NIBP times out, cuff will not inflate / hold, readings implausible.
Root cause: Internal NIBP pump or valve, a cuff / hose leak, or a pressure-transducer fault.
Symptoms: The host does not display a plugged-in module or a built-in parameter.
Root cause: Docking / connector fault or an internal board fault.
Symptoms: Blank, dim or distorted display; unresponsive touchscreen.
Root cause: Display, backlight, or controller fault.
Symptoms: Short battery run time, will not hold charge, will not power on.
Root cause: Aged battery pack or a power-supply fault.
Symptoms: Monitor not shown at the central station, admit fails, intermittent drop.
Root cause: Wired / Wi-Fi network configuration or a communication board.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable set (ECG / IBP / SpO2 / temp) | Frequent wear item | $$ | Easy |
| NIBP cuff and hose | Frequent wear item | $ | Easy |
| Internal NIBP pump / valve | Major service part | $$$ | Difficult |
| Parameter module / input board | Major service part | $$$ | Difficult |
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Display / touchscreen assembly | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
NIBP valve assembly board subject to this recall includes enhancements that are not included in the current marketing clearance:1. Anti-motion feature 2. NIBP hose blockage identification function
SmarTemp Disposable Probe covers may contain microscopic holes in the molded seam that would allow liquids or body fluids inside the cover and potentially contaminate the temperature probe.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.