Mindray · Model A5
Volume Control Ventilation (VCV). Pressure Control Ventilation (PCV) with volume guarantee. Pressure Support Ventilation (PS). Synchronized Intermittent Mandatory Ventilation in volume and pressure modes (SIMV-VC and SIMV-PC). Automatic compliance compensation and fresh gas compensation maintain accurate tidal volumes. The warmed Breathing System (35º C) virtually eliminates internal condensation. The built in caster guard design clears away cables and hoses. The economical single container absorber reduces compressible dead space volume and accepts non-proprietary prepackaged and loose fill absorbent. Integrated backup control touchpad. Deck lighting with adjustable brightness control. Two hour battery backup.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Circuit leak test fails, self-test aborts, machine not released for use.
Root cause: Breathing-circuit / absorber leak, a worn bag-arm or APL valve, a CO2-absorber canister seal, or an internal pneumatic leak.
Symptoms: Set tidal volume or pressure not delivered, PEEP not held, ventilator alarm.
Root cause: Bellows / piston-drive fault, a flow-sensor fault, or an exhalation / drive-gas valve fault.
Symptoms: FiO2 or agent value off vs a reference, hypoxic-guard alarm, agent not vaporising correctly.
Root cause: O2 sensor aging, a gas-analyser fault, a flowmeter / rotameter fault, or a vaporiser seating / back-bar leak.
Symptoms: Scavenging alarm, waste gas not evacuated.
Root cause: AGSS flow / connection or a receiver-valve fault.
Symptoms: Short battery run time, blank or frozen display.
Root cause: Aged battery or a display / power-supply fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Flow sensor(s) | Frequent wear item | $$ | Fairly easy |
| O2 sensor | Frequent wear item | $$ | Fairly easy |
| Bellows / drive assembly | Major service part | $$$ | Difficult |
| APL valve / breathing-circuit gaskets | Periodic replacement | $$ | Fairly easy |
| CO2 absorbent / canister seals | Consumable | $ | Easy |
| Battery pack | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.