Lancer · Model 810LX
• Efficient direct injection cleaning • Ideal for minimum space or budget • Flexibility with two independent wash levels and interchangeable racks • Economical and accurate dosing by two peristaltic pumps • High grade stainless steel clad on all sides for stand alone applications All Lancer washers feature these key benefits: • Fast cycles and outstanding results from powerful wash action and design • Easy to use microprocessor controls with self-diagnostic software and alarms • Flexibility of multi-level injection washing and interchangeable racks • All sizes and types of glassware cleaned with chamber configuration choices • Ergonomic and safe loading of racks on load bearing telescopic runners • Choice of preset and user-customized (PIN code protected) programs • Accurate and repeated intake of cleaning chemicals by automatic peristaltic dosing pumps • Robust high grade 304 L and 316L stainless steel construction (varies by model) • IQ/OQ protocol and validation options available • Cost effective – low life cycle costs and efficient operation uses less resources and inventory • Eco-friendly low energy, chemicals and water usage
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Pressure loss, long cycle, steam leak.
Root cause: Aged rubber gasket.
Symptoms: No steam, low temp, cycle abort.
Root cause: Failed heating element or scale buildup.
Symptoms: Temp out of tolerance, failed Bowie-Dick.
Root cause: Sensor aging.
Symptoms: Scaling, conductivity alarm.
Root cause: RO/DI system failure.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Steam generator heating element | Periodic replacement | $$$ | Difficult |
| Vacuum pump & seals | Periodic replacement | $$$ | Difficult |
| Chamber temperature sensor | Periodic replacement | $$ | Moderate |
| RO/DI feed cartridge Consumable. | Frequent wear item | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.