Lancer · Model 1400 LXP
Larger and wider chamber for greater capacity and volume with three levels of cleaning More robust control system allowing for great washer functionality Optional independent monitoring system Powerful and rapid HEPA filtered drying through injectors and in wash chamber Increased safety with on-board storage and input of chemicals Higher grade 316 L Stainless Steel Chamber Fast cycles and outstanding results from powerful wash action and design Easy to use microprocessor controls with self-diagnostic software and alarms Flexibility of multi-level injection washing and interchangeable racks All sizes and types of glassware cleaned with chamber configuration choices Ergonomic and safe loading of racks on load bearing telescopic runners Choice of preset and user-customized (PIN code protected) programs Accurate and repeated intake of cleaning chemicals by automatic peristaltic dosing pumps Robust high grade 304 L and 316L stainless steel construction (varies by model) IQ/OQ protocol and validation options available Cost effective Low life cycle costs and efficient operation uses less resources and inventory Eco-friendly low energy, chemicals and water usage.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Cycle stops with a fault code, load not processed, door locked.
Root cause: Water fill / drain fault, temperature not reached, a dosing-pump fault, or a spray-arm blockage.
Symptoms: Instruments come out soiled, spray arms not rotating.
Root cause: Clogged spray-arm nozzles, a blocked filter / strainer, or low water pressure.
Symptoms: Detergent / neutraliser / lubricant not dosed, dosing-pump alarm.
Root cause: Empty chemistry container, an air-locked or worn dosing pump, or a level sensor.
Symptoms: A0 / disinfection temperature not achieved.
Root cause: Heater element, temperature sensor, or steam supply.
Symptoms: Leaks during the cycle, door will not lock / unlock.
Root cause: Worn door gasket or a door-lock solenoid.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Spray-arm nozzles / bearings | Periodic replacement | $$ | Fairly easy |
| Chamber filter / strainer | Frequent wear item | $ | Easy |
| Peristaltic dosing pump / tubing | Periodic replacement | $$ | Moderate |
| Heater element / temperature sensor | Major service part | $$ | Moderate |
| Door gasket | Periodic replacement | $$ | Fairly easy |
| Fill / drain solenoid valve | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio