Flexible absorber change The disposable absorber used can be replaced flexibly during operation. Network capability Leon and Leon plus are network-capable and can be connected directly to popular PDMS systems. The Leon mri is also networkable. Data is transmitted via fiber optic technology without interfering with the functions of the MRI. Screen options for optimal monitoring. The 12" color TFT screen provides a clear view of patient data. The screen's touchscreen technology is designed for use in the MRI environment and is MRI compatible. In addition, a second monitor can be connected to the Leon mri for viewing data outside the MRI area.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Delivered agent concentration off, low/high agent alarm.
Root cause: Worn bellows or O-ring; agent contamination.
Symptoms: O2 low-pressure alarm, inability to ventilate.
Root cause: Pipeline disconnect or failed O2 flush valve.
Symptoms: Low minute volume, volume loss, leak alarm.
Root cause: Loose connections, torn bellows, cracked hose.
Symptoms: Rising EtCO2, patient rebreathing.
Root cause: CO2 absorbent saturated.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable - replace on schedule.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Breathing circuit / bellows | Periodic replacement | $$ | Fairly easy |
| CO2 absorber canister Consumable - replace on schedule. | Consumable | $ | Easy |
| Scavenger interface & hose | Periodic replacement | $$ | Fairly easy |
| O2 pressure regulator | Periodic replacement | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.