Kodak · Model Point-of-Care 260
Highly versatile This durable, lightweight, tabletop system expands on the powerful functionality of the KODAK Point-of-Care CR 120 System and KODAK Point-of-Care CR 140 System. It provides throughput of up to 60 plates an hour, and optional high-resolution scanning modes for even greater flexibility. Connectivity and productivity KODAK Quality Control Software, which drives the KODAK Point-of-Care CR 260 System, is designed to streamline clinical workflow and maximize productivity. With its DICOM 3.0 capabilities, the system can be seamlessly integrated with a broad variety of modality equipment, and RIS or PACS systems. Mounted on an optional wheeled Z-cart, it can be rolled into any situation where instant digital images are needed Scalability for tomorrow The KODAK Point-of-Care CR 260 System is part of the Carestream Health gallery of healthcare imaging products to satisfy your imaging needs. You can start with a small, low-cost digital capture and viewing system that sends images to a diagnostic hard-copy printer or a device that prints images on self-playing CDs. As your practice grows and workload increases, connect multiple KODAK Point-of-Care systems via your network to create a distributed CR (D-CR) solution. Connect your CR units to a HIS/RIS and PACS, or transmit images wirelessly for teleradiology applications.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Banding, streaks, ghost images, non-uniform density, dust artefacts.
Root cause: Dirty light guide / scanning optics, a worn erase-lamp array, imaging-plate wear, or contaminated transport rollers.
Symptoms: Plate jams in the reader, transport error, plate not returned.
Root cause: Worn or contaminated transport rollers / belts, or a sensor fault.
Symptoms: A faint previous image appears on the next exposure.
Root cause: Erase-lamp array degradation or a driver fault.
Symptoms: No or very noisy image, scan-error message.
Root cause: Scanning laser or the photomultiplier / detection chain.
Symptoms: Images will not display, annotate or send to PACS.
Root cause: Workstation PC or the DICOM configuration.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Transport rollers / belts | Periodic replacement | $$ | Moderate |
| Erase-lamp array | Periodic replacement | $$$ | Moderate |
| Imaging plates / cassettes | Periodic replacement | $$$ | Easy |
| Light guide / scanning optics | Major service part | $$$ | Difficult |
| Workstation PC | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio