Konica Minolta · Model Regius Model 110
Computed Radiography System The REGIUS Model 110 is a high performance single bay CR system that can process up to 76 plates per hour (14"x17"). A touch screen control station provides intuitive software tools for easy operation. Compact size and space saving design highlight the REGIUS Model 110.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Banding, streaks, ghost images, non-uniform density, dust artefacts.
Root cause: Dirty light guide / scanning optics, a worn erase-lamp array, imaging-plate wear, or contaminated transport rollers.
Symptoms: Plate jams in the reader, transport error, plate not returned.
Root cause: Worn or contaminated transport rollers / belts, or a sensor fault.
Symptoms: A faint previous image appears on the next exposure.
Root cause: Erase-lamp array degradation or a driver fault.
Symptoms: No or very noisy image, scan-error message.
Root cause: Scanning laser or the photomultiplier / detection chain.
Symptoms: Images will not display, annotate or send to PACS.
Root cause: Workstation PC or the DICOM configuration.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Transport rollers / belts | Periodic replacement | $$ | Moderate |
| Erase-lamp array | Periodic replacement | $$$ | Moderate |
| Imaging plates / cassettes | Periodic replacement | $$$ | Easy |
| Light guide / scanning optics | Major service part | $$$ | Difficult |
| Workstation PC | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
A hand grip column, even though locked in position, moved downward about 3 cm while a patient was holding the hand grip for lateral exposure. When the hand grip moved down, the patient who was holding it lost his or her balance and almost fell. This problem only occurred in Japan with one system.
If the patient demographic data is entered into the CS-1 manually, and the user fails to enter any data in one of the demographic data fields labeled as Birthday, Sex, Name or Patient Comments, etc, the RIM will populate the fields where no data has been entered with data from the previous patient.
One feature of the software on some of the devices which permits assembling three digital images into one image has occasionally had problems aligning the three images exactly.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio