06000Justifier HP sensor error (when moving to HP)
Confirm following and replace the faulty parts. • Justifier HP sensor • Sensor wire: JP41 ~ MCN3 (MDU) • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06001Justifier motor control error (when moving to HP)
Confirm following and replace the faulty parts. • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06002Justifier HP sensor error (when moving to cassette fixing position)
Confirm following and replace the faulty parts. • Justifier HP sensor • Sensor wire: JP41 ~ MCN3 (MDU) • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06003Justifier motor control error (when moving to cassette fixing position)
Confirm following and replace the faulty parts. • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06004Justifier motor control error (when moving to minute opening position)
06005Justifier motor control error (when moving to cassette fixing position 2)
06006Justifier HP sensor error (when moving to HP 2)
Confirm following and replace the faulty parts. • Justifier HP sensor • Sensor wire: JP41 ~ MCN3 (MDU) • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06007Justifier motor control error (when moving to HP 2)
Confirm following and replace the faulty parts. • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06008Justifier HP sensor error (when moving to HP 3)
Confirm following and replace the faulty parts. • Justifier HP sensor • Sensor wire: JP41 ~ MCN3 (MDU) • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06009Justifier motor control error (when moving to HP 3)
Confirm following and replace the faulty parts. • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06010Justifier HP sensor error (when moving to minute opening position)
Confirm following and replace the faulty parts. • Justifier HP sensor • Sensor wire: JP41 ~ MCN3 (MDU) • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06011Justifier motor control error (when moving to minute opening position 2)
Confirm following and replace the faulty parts. • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06015Justifier HP sensor error (when moving to HP)
Confirm following and replace the faulty parts. • Justifier HP sensor • Sensor wire: JP41 ~ MCN3 (MDU) • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06016Justifier motor control error (when moving to HP)
Confirm following and replace the faulty parts. • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06017Justifier HP sensor error (when moving to cassette fixing position)
Confirm following and replace the faulty parts. • Justifier HP sensor • Sensor wire: JP41 ~ MCN3 (MDU) • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06018Justifier motor control error (when moving to cassette fixing position)
Confirm following and replace the faulty parts. • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06019Justifier motor control error (when moving to minute opening position)
06020Justifier motor control error (when moving to cassette fixing position 2)
06021Justifier HP sensor error (when moving to HP 2)
Confirm following and replace the faulty parts. • Justifier HP sensor • Sensor wire: JP41 ~ MCN3 (MDU) • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06022Justifier motor control error (when moving to HP 2)
Confirm following and replace the faulty parts. • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06023Justifier HP sensor error (when moving to HP 3)
Confirm following and replace the faulty parts. • Justifier HP sensor • Sensor wire: JP41 ~ MCN3 (MDU) • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06024Justifier motor control error (when moving to HP 3)
Confirm following and replace the faulty parts. • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06025Justifier HP sensor error (when moving to minute opening position)
Confirm following and replace the faulty parts. • Justifier HP sensor • Sensor wire: JP41 ~ MCN3 (MDU) • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06026Justifier motor control error (when moving to minute opening position 2)
Confirm following and replace the faulty parts. • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06030Justifier HP sensor error (when moving to HP)
Confirm following and replace the faulty parts. • Justifier HP sensor • Sensor wire: JP41 ~ MCN3 (MDU) • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06031Justifier motor control error (when moving to HP)
Confirm following and replace the faulty parts. • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06032Justifier HP sensor error (when moving to cassette fixing position)
Confirm following and replace the faulty parts. • Justifier HP sensor • Sensor wire: JP41 ~ MCN3 (MDU) • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06033Justifier motor control error (when moving to cassette fixing position)
Confirm following and replace the faulty parts. • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06034Justifier motor control error (when moving to minute opening position)
06035Justifier motor control error (when moving to cassette fixing position 2)
06036Justifier HP sensor error (when moving to HP 2)
Confirm following and replace the faulty parts. • Justifier HP sensor • Sensor wire: JP41 ~ MCN3 (MDU) • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06037Justifier motor control error (when moving to HP 2)
Confirm following and replace the faulty parts. • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06038Justifier HP sensor error (when moving to HP 3)
Confirm following and replace the faulty parts. • Justifier HP sensor • Sensor wire: JP41 ~ MCN3 (MDU) • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06039Justifier motor control error (when moving to HP 3)
Confirm following and replace the faulty parts. • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06040Justifier HP sensor error (when moving to minute opening position)
Confirm following and replace the faulty parts. • Justifier HP sensor • Sensor wire: JP41 ~ MCN3 (MDU) • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06041Justifier motor control error (when moving to minute opening position 2)
Confirm following and replace the faulty parts. • Justifier motor • Motor wire: JJ50 ~ MCN5 (MDU) • MDU
06050Receiver HP sensor error (when moving to HP)
Confirm following and replace the faulty parts. • Receiver HP sensor • Sensor wire: JP43 ~ MCN3 (MDU) • Receiver motor • Motor wire: JP51 ~ MCN5 (MDU) • MDU
06051Receiver motor control error (when moving to HP)
Confirm following and replace the faulty parts. • Receiver motor • Motor wire: JP51 ~ MCN5 (MDU) • MDU
06052Receiver HP sensor error (when moving to first falling position)
Confirm following and replace the faulty parts. • Receiver HP sensor • Sensor wire: JP43 ~ MCN3 (MDU) • Receiver motor • Motor wire: JP51 ~ MCN5 (MDU) • MDU
06053Receiver motor control error (when moving to first falling position)
Confirm following and replace the faulty parts. • Receiver motor • Motor wire: JP51 ~ MCN5 (MDU) • MDU
06054Receiver motor control error (when moving to second falling position)
Confirm following and replace the faulty parts. • Receiver motor • Motor wire: JP51 ~ MCN5 (MDU) • MDU
06055Receiver motor control error (when moving slightly down at the second falling position)
Confirm following and replace the faulty parts. • Receiver motor • Motor wire: JP51 ~ MCN5 (MDU) • MDU
06056Receiver motor control error (when moving to first falling position 2)
Confirm following and replace the faulty parts. • Receiver motor • Motor wire: JP51 ~ MCN5 (MDU) • MDU
06057Receiver HP sensor error (when moving to HP 2)
Confirm following and replace the faulty parts. • Receiver HP sensor • Sensor wire: JP43 ~ MCN3 (MDU) • Receiver motor • Motor wire: JP51 ~ MCN5 (MDU) • MDU
06058Receiver motor control error (when moving to HP 2)
Confirm following and replace the faulty parts. • Receiver motor • Motor wire: JP51 ~ MCN5 (MDU) • MDU
06059Receiver HP sensor error (when moving to HP 3)
Confirm following and replace the faulty parts. • Receiver HP sensor • Sensor wire: JP43 ~ MCN3 (MDU) • Receiver motor • Motor wire: JP51 ~ MCN5 (MDU) • MDU
06060Receiver motor control error (when moving to HP 3)
Confirm following and replace the faulty parts. • Receiver motor • Motor wire: JP51 ~ MCN5 (MDU) • MDU
06061Receiver HP sensor error (when moving to second falling position)
Confirm following and replace the faulty parts. • Receiver HP sensor • Sensor wire: JP43 ~ MCN3 (MDU) • Receiver motor • Motor wire: JJ51 ~ MCN5 (MDU) • MDU
06062Receiver motor control error (when moving to second falling position 2)
Confirm following and replace the faulty parts. • Receiver motor • Motor wire: JJ51 ~ MCN5 (MDU) • MDU
06100Transporter read sensor error (when moving to read position)
1. Confirm that there are no foreign objects between the push plate and the frame, and remove if there are any. 2. Confirm if there is enough grease on the drive shaft, nut, or transporter motor unit, and apply grease if it is not enough. 3. Confirm following and replace the faulty parts. • Tumbler (if the spring is not broken) • Transporter read sensor • Sensor wire: JP47 ~ MCN4 (MDU) • Transporter motor • Motor wire: JP52 ~ MCN6 (MDU) • MDU
06101Transporter motor control error (when moving to read position)
1. Confirm the ground. (It may cause noise) 2. Confirm that the antistatic brush right under the insertion slot has not fallen off. 3. Confirm following and replace the faulty parts. • Ground wire of transporter unit (if it is not broken) • Transporter motor • Motor wire: JP52 ~