Kodak · Model Point-of-Care 120
Highly versatile The POINT-OF-CARE CR 120 System can be configured for most clinical applications, including specialties practices such as chiropractors, podiatrists, and orthopaedists. A simple-to-use KODAK Quality Control Scanner interface designed for full DICOM connectivity allows the CR 120 to capture high-quality x-ray images of any body part. All imaging parameters are optimized, resulting in digital images that can be enhanced, enlarged, duplicated, and sent to any location in seconds with no loss of resolution. Images can be printed or archived locally on CDs or DVDs. POINT-OF-CARE productivity Mounted on an optional wheeled Z-cart or placed on a tabletop, the KODAK POINT-OF-CARE CR 120 System can be used in virtually any location. With the Z-cart, this system can be rolled into any situation where nearly instant digital images are needed. Mobile x-ray The Point-of Care CR 120 System can provide a complete mobile digital imaging solution on wheels. Mount the system in a van or truck and combine it with reusable KODAK flexible phosphor screens and cassettes, a rugged laptop computer, and portable x-ray equipment. Then drive anywhere x-ray exams are needed. KODAK POINT-OF-CARE mobile solutions serve nursing facilities, prisons, forensic institutions, employee screening needs, and more throughout the world.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Banding, streaks, ghost images, non-uniform density, dust artefacts.
Root cause: Dirty light guide / scanning optics, a worn erase-lamp array, imaging-plate wear, or contaminated transport rollers.
Symptoms: Plate jams in the reader, transport error, plate not returned.
Root cause: Worn or contaminated transport rollers / belts, or a sensor fault.
Symptoms: A faint previous image appears on the next exposure.
Root cause: Erase-lamp array degradation or a driver fault.
Symptoms: No or very noisy image, scan-error message.
Root cause: Scanning laser or the photomultiplier / detection chain.
Symptoms: Images will not display, annotate or send to PACS.
Root cause: Workstation PC or the DICOM configuration.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Transport rollers / belts | Periodic replacement | $$ | Moderate |
| Erase-lamp array | Periodic replacement | $$$ | Moderate |
| Imaging plates / cassettes | Periodic replacement | $$$ | Easy |
| Light guide / scanning optics | Major service part | $$$ | Difficult |
| Workstation PC | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.