HP · Model CodeMaster 100
12 lead defibrillator 3 configurable pairs of high and low heart rate alarm limits from 20 to 280 bpm Durable design and extra-rugged materials made to withstand everyday bumps and scrapes Built-in pediatric paddles immediately available under slide-off adult electrodes SYNC message appears on monitor and is annoted periodically on recorder while in synchronous mode. An audible beep sounds with each detected R-wave, while a marker on the monitor and sync designator on the recorder strip indicate the discharge point. Armed Indicator: Charge done tone, charge done light on apex paddle and on defibrillator front panel and available energy indicated on display
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Charge fails or sluggish, low-energy delivery.
Root cause: Capacitor aging / leakage.
Symptoms: Short charge-hold, low-battery on readiness test.
Root cause: Aged battery pack.
Symptoms: No shock delivered, pad error.
Root cause: Worn paddles or expired pads.
Symptoms: Leakage alarm, self-test fail.
Root cause: Failed isolation components.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Expiration-dated consumable.
Critical, keep charged spares.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Defibrillation electrodes / pads Expiration-dated consumable. | Consumable | $$ | Easy |
| Charge relay / contactor | Periodic replacement | $$$ | Difficult |
| Patient interface cable | Periodic replacement | $$ | Fairly easy |
| Rechargeable battery pack Critical, keep charged spares. | Periodic replacement | $$$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.