HP · Model 43100A
A built-in annotating recorder and internal battery The HP 43100/110A meets current ACLS standards Uses microprocessor technology to control and monitor system operation, which allows the unit to perform a diagnostic routine Produces up to 360 joules Charge done tone, and charge done lamp on apex paddle to indicate that it is armed 3-color LED bar graph array on STERNUM paddle indicates quality of debrillator paddle contact before discharge
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Charge fails or sluggish, low-energy delivery.
Root cause: Capacitor aging / leakage.
Symptoms: Short charge-hold, low-battery on readiness test.
Root cause: Aged battery pack.
Symptoms: Won't charge or discharges unexpectedly.
Root cause: Relay/contactor wear.
Symptoms: Leakage alarm, self-test fail.
Root cause: Failed isolation components.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Expiration-dated consumable.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| HV energy storage capacitor | Major service part | $$$$ | Expert-level |
| Defibrillation electrodes / pads Expiration-dated consumable. | Consumable | $$ | Easy |
| Charge relay / contactor | Periodic replacement | $$$ | Difficult |
| Patient interface cable | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →