Hospira · Model Plum 360
• Air Matters - the Plum 360 utilizes the unique PlumSet™ cassette that features air management that doesn’t require disconnecting from the patient • Remove air faster - use the pump to automatically remove air through the cassette’s secondary administration port without disconnecting the line from the patient • Reduce therapy interruptions - capture 1 mL of air in the cassette’s air trap – an industry exclusive feature – before the alarm is activated • Reduce risk of contamination - Help eliminate the potential for contaminants by keeping the patient line connected • Secondary Delivery is of Primary Importance - Hospira’s Plum infusion systems are unique in the way they connect the secondary line directly to the PlumSet cassette, so the pump can differentiate between primary and secondary infusion lines and deliver each as intended • Optimize Safety and Compliance at the Point of Care - when programming your Plum 360 infusion system with Hospira MedNet™ software, you automatically start in the drug library • EMR Ready, Today and Tomorrow - when you and your electronic medical record (EMR) partner are ready for integration, your Plum 360 infusion system with Hospira MedNet safety software is ready too. Whether it’s today, or any time after, Hospira’s IV • Clinical Integration solution will connect your Plum 360 system to the medication order, patient information and EMR
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Occlusion alarm, no flow, upstream pressure high.
Root cause: Kinked set, clamped line, patient vein issue.
Symptoms: Air-in-line alarm, infusion paused.
Root cause: Air in set, improperly primed tubing.
Symptoms: Uncontrolled flow when door open.
Root cause: Worn roller or failed free-flow valve.
Symptoms: Set not recognized, door-open error.
Root cause: Dirty or failed door microswitch.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable; per patient.
Common R&R on high-mileage pumps.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Door microswitch | Periodic replacement | $ | Fairly easy |
| Occlusion sensor assembly | Periodic replacement | $$ | Moderate |
| IV administration set Consumable; per patient. | Consumable | $ | Easy |
| Peristaltic roller / pump mechanism Common R&R on high-mileage pumps. | Periodic replacement | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.