HealthySole · Model UVC Germicidal Disinfection
Introducing groundbreaking use of UVC technology that eliminates the dangerous organisms that cause healthcare-associated infections (HAI's). Add a significant layer of defense to your existing infection control/prevention program. Once implemented, this product will reduce the rate of airborne, horizontal and cross contamination - without labor costs!
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: No power, indicators dark.
Root cause: AC power not present (cord not seated, wall receptacle dead, tripped breaker) or a blown AC fuse in the snap-in holder between the ON/OFF switch and the cord socket.
Symptoms: System powers on but the UVC shutters stay closed, no exposure.
Root cause: Safety interlock not made (base cover / dust shield not seated after service), the interlock cable unplugged, or an IR foot/heel sensor cable disconnected.
Symptoms: Shutters open but no UVC output, treatment time counts with no disinfection.
Root cause: UVC lamp(s) at end of life, a lamp not seated, or a lamp-driver / ballast fault.
Symptoms: Cycle does not start when a user steps on, or ends early.
Root cause: Toe or heel IR sensor obstructed, dirty, or its cable unplugged from the system board.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Handle with gloves; re-seat the safety-interlock cable after replacement.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| UVC lamps Handle with gloves; re-seat the safety-interlock cable after replacement. | Frequent wear item | $$ | Moderate |
| AC fuses (snap-in pair) | Major service part | $ | Fairly easy |
| IR toe / heel sensors | Periodic replacement | $$ | Moderate |
| Dust shield | Frequent wear item | $ | Fairly easy |
| Lamp driver / ballast | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.