60-inch walking surface Two stopping modes Standard full handrail set Standard emergency stop switch Starts from 0 mph 0 to 13.5 mph speed range Weight capacity of 450 lbs. 0 to 25 percent grade range.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Walking belt slips under load, tracks off-centre, jerky or noisy.
Root cause: Worn or stretched belt, a worn deck, loss of drive-belt tension, or worn rollers.
Symptoms: Console dead, belt will not move, incline will not change, error code.
Root cause: Breaker / power, a drive-motor or controller fault, a lift-motor fault, or a console-cable fault.
Symptoms: Actual speed / grade does not match the display or protocol.
Root cause: Speed-sensor or incline-potentiometer / limit-switch fault - needs calibration.
Symptoms: Motor cuts out under load, burning smell, controller error.
Root cause: Worn motor brushes / bearings or a drive-controller fault.
Symptoms: E-stop or safety key does not stop the belt, or stops it unexpectedly.
Root cause: Failed e-stop switch / magnet or its wiring - a safety function.
Symptoms: Intermittent signal/power.
Root cause: Repeated flexing.
Symptoms: Short runtime, low-battery alarm.
Root cause: Aged battery.
Symptoms: Loose device, vibration damage.
Root cause: Docking latch wear.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Critical for transport; spares mandatory.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Walking belt & deck | Periodic replacement | $$ | Moderate |
| Drive belt | Periodic replacement | $ | Moderate |
| Drive motor / controller | Major service part | $$$ | Difficult |
| Incline (lift) motor / potentiometer | Major service part | $$ | Moderate |
| Roller / bearing set | Major service part | $$ | Moderate |
| Patient cable set | Periodic replacement | $$ | Easy |
| Rechargeable battery pack Critical for transport; spares mandatory. | Frequent wear item | $$$ | Fairly easy |
| Docking cradle / latch | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
GE Healthcare has recently become aware of a potential safety issue due to uncontrolled motion of the walking belt on your T2100 Treadmill.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio